Investigation of the mixing efficiency of a chaotic micromixer using thermal lens spectrometry.
Ghaleb, Khalil Abbas; Stephan, Khaled; Pittet, Patrick; Ferrigno, Rosaria; Georges, Joseph.
Afiliação
  • Ghaleb KA; Laboratoire des Sciences Analytiques, UMR 5180, Bât. CPE-308D, Domaine Scientifique de la Doua, Université Claude Bernard-Lyon 1, 69622 Villeurbanne Cedex, France.
Appl Spectrosc ; 60(5): 564-7, 2006 May.
Article em En | MEDLINE | ID: mdl-16756709
ABSTRACT
This work investigates the efficiency of a chaotic micromixer using thermal lens spectrometry. The outlet of the mixing device was connected to a thermal lens detection head integrating the probe beam optical fibers and the sample capillary. The chaotic micromixer consisted of a Y-shaped poly(dimethylsiloxane) (PDMS) microchip in which ribbed herringbone microstructures were etched on the floor of the main channel. Due to the solvent composition dependence of the thermal lens response, the photothermal method was shown to be highly sensitive to nonhomogeneous mixing compared to fluorescence detection. The apparatus was applied to the determination of Fe2+ with 1,10-phenanthroline using flow injection analysis; a limit of detection of 11 microg L(-1) of iron was obtained.