Your browser doesn't support javascript.
loading
A new device for the treatment of thromboembolic strokes.
Pearce, Gillian; Patrick, John H; Perkinson, Neil D.
Afiliação
  • Pearce G; Robert Jones and Agnus Hunt, Oswestry, United Kingdom. G.Pearce@wlv.ac.uk
J Stroke Cerebrovasc Dis ; 16(4): 167-72, 2007.
Article em En | MEDLINE | ID: mdl-17689413
ABSTRACT
We present data concerning the design and testing of a new clot-sucking device for possible use in the extraction of blood clots that arise during thromboembolic strokes. Our data indicate that a jelly-like mass of 0.02 g (of an artificial clot) can be removed from plastic tubing using this vortex-creating device, which has an optimal diameter of 1 mm, in conjunction with a 6F arterial catheter that is 110-cm long and using applied suction pressures of about 50 cm Hg.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Trombectomia / Embolia Intracraniana / Trombose Intracraniana Limite: Humans Idioma: En Revista: J Stroke Cerebrovasc Dis Assunto da revista: ANGIOLOGIA / CEREBRO Ano de publicação: 2007 Tipo de documento: Article País de afiliação: Reino Unido
Buscar no Google
Base de dados: MEDLINE Assunto principal: Trombectomia / Embolia Intracraniana / Trombose Intracraniana Limite: Humans Idioma: En Revista: J Stroke Cerebrovasc Dis Assunto da revista: ANGIOLOGIA / CEREBRO Ano de publicação: 2007 Tipo de documento: Article País de afiliação: Reino Unido