Conducting polymer based active catheter for minimally invasive interventions inside arteries.
Annu Int Conf IEEE Eng Med Biol Soc
; 2008: 2063-6, 2008.
Article
em En
| MEDLINE
| ID: mdl-19163101
ABSTRACT
An active catheter intended for controllable intravascular maneuvers is presented and initial experimental results are shown. A commercial catheter is coated with polypyrrole and laser micromachined into electrodes, which are electrochemically activated, leading to bending of the catheter. The catheter's electro-chemo-mechanical properties are theoretically modeled to design the first prototype device, and used to predict an optimal polypyrrole thickness for the desired degree of bending within approximately 30 seconds. We compared the experimental result of catheter bending to the theoretical model with estimated electrochemical strain, showing reasonable agreement. Finally, we used the model to design an encapsulated catheter with polypyrrole actuation for improved intravascular compatibility and performance.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Artérias
/
Cateterismo
/
Procedimentos Cirúrgicos Minimamente Invasivos
Tipo de estudo:
Prognostic_studies
Idioma:
En
Revista:
Annu Int Conf IEEE Eng Med Biol Soc
Ano de publicação:
2008
Tipo de documento:
Article