Your browser doesn't support javascript.
loading
LC/MS/MS Analysis of N-Terminal Protein Adducts with Improved Sensitivity: A Comparison of Selected Edman Isothiocyanate Reagents.
Rydberg, Per; von Stedingk, Hans; Magnér, Jörgen; Björklund, Jonas.
Afiliação
  • Rydberg P; Arrhenius Laboratory, Department of Environmental Chemistry, Stockholm University, 106 91 Stockholm, Sweden.
Int J Anal Chem ; 2009: 153472, 2009.
Article em En | MEDLINE | ID: mdl-20107558
ABSTRACT
This study provides a basis for a new and straightforward method for LC/MS/MS-based screening of N-terminal protein adducts. This procedure is denoted the "FIRE procedure" as fluorescein isothiocyanate (FITC) gave superior sensitivity by LC/MS/MS when measuring adducts (R) of electrophilic compounds with a modified Edman procedure. The principles of the FIRE-procedure are that adducts to N-terminal amino acids selectively are detached and measured from of proteins after derivatisation by isothiocyanate Edman reagents. In this study, FITC, 4-N,N-dimethylaminoazobenzene 4'-isothiocyanate (DABITC) and 4-dimethylamino-1-naphthyl isothiocyanate (DNITC) were used to synthesize thiohydantoin analytes from valine and N-methylvaline. The sensitivity by LC/MS/MS was enhanced by up to three orders of magnitude as compared to phenyl isothiocyanate and higher as compared to pentafluorophenyl isothiocyanate. The FITC reagent will enable measurements of low background adduct levels. Synthesized analytes were characterised with, for example, (1)H NMR, (13)C NMR, LC/MS/MS, and UV.

Texto completo: 1 Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Revista: Int J Anal Chem Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Suécia

Texto completo: 1 Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Revista: Int J Anal Chem Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Suécia