The characteristics of dental X-ray fluoroscopic equipment 'DreamRay 60F'.
Shin, Gwi-Soon; Lee, Bo-Ram; Chang, Kyung-Hwan; Lee, In-Ja; Lee, Wonho; Choi, Jong-Hak; Kim, You-Hyun.
Afiliação
  • Shin GS; Medical Physics Laboratory, Department of Radiological Technology, College of Health Science, Korea University, Seoul, Republic of Korea.
Radiat Prot Dosimetry ; 140(4): 362-8, 2010 Aug.
Article em En | MEDLINE | ID: mdl-20554581
This study examined the characteristics of the dental X-ray fluoroscopic equipment, 'DreamRay 60F', which was recently developed in Korea. The output linearity, output reproducibility, half-value layer (HVL), leakage radiation and scattered radiation were measured using an ionisation chamber. The surface dose equivalent rate and estimated dose equivalent of the operator were also calculated. The output linearity was 0.0015-0.0175 and the coefficient of variation for the output reproducibility was 0.0013-0.0074. The experimental HVL was 2.1 mm Al, and the leakage dose rate at 100 cm from the X-ray focus ranged from 2.70 to 19.66 microGy h (-1) depending on the direction. The scattered radiation doses differed significantly (1.7-16.8 times) depending on the distance and direction. If an operator is exposed for 10 min per procedure, 5 procedures a day at 5 days a week, he/she sitting at a 90 degrees direction will receive an annual dose equivalent of 13.0 mSv (at 30 cm) and 63.7 mSv (at 50 cm) in the trunk and face surface, respectively.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doses de Radiação / Radiometria / Fluoroscopia / Radiografia Dentária Tipo de estudo: Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Radiat Prot Dosimetry Ano de publicação: 2010 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doses de Radiação / Radiometria / Fluoroscopia / Radiografia Dentária Tipo de estudo: Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Radiat Prot Dosimetry Ano de publicação: 2010 Tipo de documento: Article