Your browser doesn't support javascript.
loading
Spectroscopic sensitive polarimeter for biomedical applications.
Ramella-Roman, Jessica C; Nayak, Amritha; Prahl, Scott A.
Afiliação
  • Ramella-Roman JC; Biomedical Engineering Department, The Catholic University of America, Michigan Avenue NE, Washington, DC 20064, USA. ramella@cua.edu
J Biomed Opt ; 16(4): 047001, 2011 Apr.
Article em En | MEDLINE | ID: mdl-21529091
We present the design and calibration of a spectroscopic sensitive polarimeter. The polarimeter can measure the full Stokes vector in the wavelength range 550 to 750 nm with 1-nm resolution and consists of a fiber-based spectrophotometer, a white light emitting diode light source, two liquid crystal retarders, and one polarizer. Calibration of the system is achieved with a scheme that does not require knowledge of the polarizing elements' orientation or retardation. Six intensity spectra are required to calculate the full spectrum Stokes vector. Error in the polarimeter is less than 5%. We report the Stokes vectors for light transmitted through nonscattering polarizing elements as well as a measurement of the depolarizing properties of chicken muscle at several wavelengths.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Análise Espectral / Óptica e Fotônica Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Revista: J Biomed Opt Assunto da revista: ENGENHARIA BIOMEDICA / OFTALMOLOGIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Análise Espectral / Óptica e Fotônica Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Revista: J Biomed Opt Assunto da revista: ENGENHARIA BIOMEDICA / OFTALMOLOGIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Estados Unidos