Thiocyanate and nitrite analysis using miniaturised isotachophoresis on a planar polymer chip.
Analyst
; 136(15): 3170-6, 2011 Aug 07.
Article
em En
| MEDLINE
| ID: mdl-21674078
A new method has been developed to improve the determination of thiocyanate using isotachophoresis. This method uses complexation with copper(II) as a mechanism for improving the separation of thiocyanate from chlorate and perchlorate. By using a pH of 3.25 the method can also be used to analyse nitrite. Separations were carried out using a miniaturised poly(methyl methacrylate) (PMMA) separation device. Linearity was observed from 1.25 to 75 mg dm(-3) with a correlation coefficient of 0.998 for both thiocyanate and nitrite. Limits of detection for these two species were calculated to be 0.8 mg dm(-3) and 0.9 mg dm(-3) respectively. The method was successfully applied to the analysis of these anions in a range of samples including explosive residues.
Texto completo:
1
Base de dados:
MEDLINE
Idioma:
En
Revista:
Analyst
Ano de publicação:
2011
Tipo de documento:
Article