Your browser doesn't support javascript.
loading
Analytical performance and clinical concordance of the cancer biomarkers CA 15-3, CA 19-9, CA 125 II, Carcinoembryonic Antigen, and Alpha-Fetoprotein on the Dimension Vista® System.
Christenson, Robert H; Cervelli, Denise R; Sterner, Jamie; Bachmann, Lauren M; Rebuck, Heather; Gray, Jeffery; Kelley, Walter E.
Afiliação
  • Christenson RH; University of Maryland School of Medicine, Baltimore, MD, USA. Electronic address: RCHRISTENSON@umm.edu.
  • Cervelli DR; Siemens Healthcare Diagnostics, Newark, DE, USA.
  • Sterner J; Siemens Healthcare Diagnostics, Newark, DE, USA.
  • Bachmann LM; VCU Health System, Richmond, VA, USA.
  • Rebuck H; University of Maryland School of Medicine, Baltimore, MD, USA.
  • Gray J; University of Maryland School of Medicine, Baltimore, MD, USA.
  • Kelley WE; University of Maryland Medical Center, Baltimore, MD, USA.
Clin Biochem ; 44(13): 1128-1136, 2011 Sep.
Article em En | MEDLINE | ID: mdl-21704611
ABSTRACT

OBJECTIVES:

We examined analytical characteristics of new CA 15-3, CA 19-9, CA 125 II, Carcinoembryonic Antigen (CEA), and Alpha-Fetoprotein (AFP) assays on the Dimension Vista® System. DESIGN AND

METHODS:

Imprecision studies used CLSI-EP5-A2, Limit of Blank and Limit of Detection used CLSI-EP17 and measurement ranges were determined. Method comparisons were evaluated with Passing-Bablok, least-squares regression and residual plots. Reference intervals were determined and valid specimen types, lot-to-lot variability and sample storage stability were defined. Clinical monitoring patterns for each tumor marker in patients were examined.

RESULTS:

Reproducibility for each method was <6.5%. Limits of Blank and Detection were low. Comparisons between methods showed slopes ranging from 0.89 to 1.32 with low y-intercepts and scatter. Minimal lot-to-lot variability was documented; serum/plasma specimens provide valid results; sample stability at -70°C was >9months. Clinical monitoring patterns correlated with established methods in >89% of cases.

CONCLUSIONS:

Measurement of CA 15-3, CA 19-9, CA 125 II, CEA and AFP on the Dimension Vista® System is an attractive alternative.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Antígenos Glicosídicos Associados a Tumores / Alfa-Fetoproteínas / Biomarcadores Tumorais / Técnicas e Procedimentos Diagnósticos / Neoplasias Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: En Revista: Clin Biochem Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Antígenos Glicosídicos Associados a Tumores / Alfa-Fetoproteínas / Biomarcadores Tumorais / Técnicas e Procedimentos Diagnósticos / Neoplasias Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: En Revista: Clin Biochem Ano de publicação: 2011 Tipo de documento: Article