Improving fixation strength of pedicle screw by microarc oxidation treatment: an experimental study of osteoporotic spine in sheep.
J Orthop Res
; 30(8): 1296-303, 2012 Aug.
Article
em En
| MEDLINE
| ID: mdl-22234960
Failure of fixation caused by loosening of pedicle screws in osteoporosis is a problem in spinal surgery. We compared the in vivo fixation strength between pedicle screws treated with microarc oxidation (MAO) and untreated screws in an osteoporotic model of ovariectomized sheep. The MAO treated and untreated screws were placed in lumbar vertebral bodies. After 3 months of implantation, biomechanical tests, micro-CT analysis, and histological observations were conducted to examine the performance of the two groups. At time 0, no significant difference was found between the two groups in biomechanical tests (p > 0.05); 3 months later, higher pull-out strength and load with less displacement were detected in the MAO-treated group (p < 0.05). Micro-CT analysis showed that the tissue mineral density, bone volume fraction, trabecular thickness, and trabecular number in the MAO-treated group were all higher than those in untreated group (p < 0.05), and trabecular spacing was smaller (p < 0.05). Histologically, the bone-implant interface in the MAO-treated group was better than that in untreated group (p < 0.05). In conclusion, pedicle screws with a bioactive surface treated by MAO can improve screw fixation strength in osteoporotic spines in sheep.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Osteoporose
/
Parafusos Ósseos
/
Fixadores Internos
Tipo de estudo:
Prognostic_studies
Limite:
Animals
Idioma:
En
Revista:
J Orthop Res
Ano de publicação:
2012
Tipo de documento:
Article
País de afiliação:
China