Recurrent pericardial effusion caused by pacemaker lead perforation and warfarin therapy at seven years after implantation.
Nakanishi, Hiroyuki; Kashiwase, Kazunori; Nishio, Mayu; Wada, Mitsuru; Hirata, Akio; Ueda, Yasunori.
Afiliação
  • Nakanishi H; Cardiovascular Division, Osaka Police Hospital, 10-31 Kitayama-tyo, Tennohji-ku, Osaka 543-0035, Japan.
Europace ; 14(2): 297, 2012 Feb.
Article em En | MEDLINE | ID: mdl-22266846
ABSTRACT
A 66-year-old man was implanted with a pacemaker. Seven years after implantation he was admitted due to cardiogenic cerebral embolism and warfarin therapy was introduced. After that, he suffered recurrent pericardial effusion for unexplained reasons. An exploratory thoracotomy revealed that the screw of the atrial lead had penetrated through the right auricular appendage wall.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Derrame Pericárdico / Varfarina / Ferimentos Penetrantes / Apêndice Atrial / Eletrodos Implantados Limite: Aged / Humans / Male Idioma: En Revista: Europace Assunto da revista: CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Derrame Pericárdico / Varfarina / Ferimentos Penetrantes / Apêndice Atrial / Eletrodos Implantados Limite: Aged / Humans / Male Idioma: En Revista: Europace Assunto da revista: CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Japão