Your browser doesn't support javascript.
loading
[Experimental study of safety in application of the Iris Medical IQ 810 diode laser in clinical treatment of age-related macular degeneration].
Patol Fiziol Eksp Ter ; (3): 16-20, 2011.
Article em Ru | MEDLINE | ID: mdl-22279734
ABSTRACT
During the study of diode laser radiation effect in micropulse mode on culture cells of human retinal pigment epithelium it was revealed that the quota of dead cells was a minimum. Besides, a certain conformity between dead cells quota and parameter characteristics of laser radiation. Based on the performed experimental study it was revealed that for a work using the Iris Medical IQ 810 diode laser in the micropulse mode following parameters duration of pulse set--300ms, duration of function--9.1%, power--750mW are safe for retinal pigment epithelium cells. Rationales of safety in application of the infrared diode laser radiation in micropulse mode in clinic for treatment of age-related macular degeneration (AMD) exampled by cell culture of human retinal pigment epithelium.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Epitélio Pigmentado da Retina / Lasers / Degeneração Macular Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: Ru Revista: Patol Fiziol Eksp Ter Ano de publicação: 2011 Tipo de documento: Article
Buscar no Google