Validation of the Better Care® system to detect ineffective efforts during expiration in mechanically ventilated patients: a pilot study.
Blanch, Lluis; Sales, Bernat; Montanya, Jaume; Lucangelo, Umberto; Garcia-Esquirol, Oscar; Villagra, Ana; Chacon, Encarna; Estruga, Anna; Borelli, Massimo; Burgueño, Ma Jose; Oliva, Joan C; Fernandez, Rafael; Villar, Jesus; Kacmarek, Robert; Murias, Gastón.
Afiliação
  • Blanch L; Critical Care Center, Hospital de Sabadell, Corporacio Sanitaria Universitària Parc Tauli, Universitat Autònoma de Barcelona, Parc Taulí 1, 08208 Sabadell, Spain. lblanch@tauli.cat
Intensive Care Med ; 38(5): 772-80, 2012 May.
Article em En | MEDLINE | ID: mdl-22297667
ABSTRACT

PURPOSE:

Ineffective respiratory efforts during expiration (IEE) are a problem during mechanical ventilation (MV). The goal of this study is to validate mathematical algorithms that automatically detect IEE in a computerized (Better Care®) system that obtains and processes data from intensive care unit (ICU) ventilators in real time.

METHODS:

The Better Care® system, integrated with ICU health information systems, synchronizes and processes data from bedside technology. Algorithms were developed to analyze airflow waveforms during expiration to determine IEE. Data from 2,608,800 breaths from eight patients were recorded. From these breaths 1,024 were randomly selected. Five experts independently analyzed the selected breaths and classified them as IEE or not IEE. Better Care® evaluated the same 1,024 breaths and assigned a score to each one. The IEE score cutoff point was determined based on the experts' analysis. The IEE algorithm was subsequently validated using the electrical activity of the diaphragm (EAdi) signal to analyze 9,600 breaths in eight additional patients.

RESULTS:

Optimal sensitivity and specificity were achieved by setting the cutoff point for IEE by Better Care® at 42%. A score >42% was classified as an IEE with 91.5% sensitivity, 91.7% specificity, 80.3% positive predictive value (PPV), 96.7% negative predictive value (NPV), and 79.7% Kappa index [confidence interval (CI) (95%) = (75.6%; 83.8%)]. Compared with the EAdi, the IEE algorithm had 65.2% sensitivity, 99.3% specificity, 90.8% PPV, 96.5% NPV, and 73.9% Kappa index [CI (95%) = (71.3%; 76.3%)].

CONCLUSIONS:

In this pilot, Better Care® classified breaths as IEE in close agreement with experts and the EAdi signal.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Respiração Artificial / Expiração / Monitorização Fisiológica Tipo de estudo: Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adolescent / Aged / Aged80 / Female / Humans / Male / Middle aged País/Região como assunto: Europa Idioma: En Revista: Intensive Care Med Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Espanha

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Respiração Artificial / Expiração / Monitorização Fisiológica Tipo de estudo: Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adolescent / Aged / Aged80 / Female / Humans / Male / Middle aged País/Região como assunto: Europa Idioma: En Revista: Intensive Care Med Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Espanha