  • Sievert H; Johann Wolfgang Goethe University, Frankfurt, Germany; Katharinen Hospital, Stuttgart, Germany; Cardiovascular Center, Frankfurt, Germany; Sankt Katharinen Hospital, Frankfurt, Germany; Department Internal Medicine, Cardiology and Vascular Medicine, St. Josefs-Hospital, Wiesbaden, Germany.
Catheter Cardiovasc Interv ; 83(7): 1135-43, 2014 Jun 01.
Article em En | MEDLINE | ID: mdl-22815250
ABSTRACT

OBJECTIVE:

The Coherex-EU Study evaluated the safety and efficacy of PFO closure utilizing novel in-tunnel PFO closure devices.

BACKGROUND:

Transcatheter closure of patent foramen ovale (PFO) followed the development of transcatheter closure devices designed to patch atrial septal defects (ASDs). The Coherex FlatStent™ and FlatStent™ EF devices were designed specifically to treat PFO anatomy.

METHODS:

A total of 95 patients with a clinical indication for PFO closure were enrolled in a prospective, multicenter first in man study at six clinical sites. Thirty-six patients received the first-generation FlatStent study device, and 57 patients received the second-generation FlatStent EF study device, which was modified based on clinical experience during the first 38 cases. Two patients enrolled to receive the first generation did not receive a device.

RESULTS:

At 6 months post-procedure, 45% (17/38) of the intention-to-treat (ITT) cohort receiving the first-generation FlatStent device had complete closure, 26% (10/38) had a trivial residual shunt, and 29% (11/38) had a moderate to large residual shunt. In the ITT cohort receiving the second-generation FlatStent EF device, 76% (43/57) had complete closure, 12% (7/57) had a trivial shunt, and 12% had a moderate to large shunt. Five major adverse events occurred, all without sequelae.

CONCLUSION:

This initial study of the Coherex FlatStent/FlatStent EF PFO Closure System demonstrated the potential for in-tunnel PFO closure. The in-tunnel Coherex FlatStent EF may offer an alternative to septal repair devices for PFO closure in appropriately selected patients; however, further investigation will be necessary to establish the best use of this device.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Cateterismo Cardíaco / Stents / Forame Oval Patente / Dispositivo para Oclusão Septal / Procedimentos Cirúrgicos Cardíacos Tipo de estudo: Clinical_trials / Diagnostic_studies / Observational_studies / Prognostic_studies Limite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Catheter Cardiovasc Interv Assunto da revista: CARDIOLOGIA Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Cateterismo Cardíaco / Stents / Forame Oval Patente / Dispositivo para Oclusão Septal / Procedimentos Cirúrgicos Cardíacos Tipo de estudo: Clinical_trials / Diagnostic_studies / Observational_studies / Prognostic_studies Limite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Catheter Cardiovasc Interv Assunto da revista: CARDIOLOGIA Ano de publicação: 2014 Tipo de documento: Article País de afiliação: Alemanha