Reduction in inappropriate therapy and mortality through ICD programming.
Moss, Arthur J; Schuger, Claudio; Beck, Christopher A; Brown, Mary W; Cannom, David S; Daubert, James P; Estes, N A Mark; Greenberg, Henry; Hall, W Jackson; Huang, David T; Kautzner, Josef; Klein, Helmut; McNitt, Scott; Olshansky, Brian; Shoda, Morio; Wilber, David; Zareba, Wojciech.
Afiliação
  • Moss AJ; Department of Medicine, University of Rochester Medical Center, Rochester, NY 14642-0653, USA. heartajm@heart.rochester.edu
N Engl J Med ; 367(24): 2275-83, 2012 Dec 13.
Article em En | MEDLINE | ID: mdl-23131066
ABSTRACT

BACKGROUND:

The implantable cardioverter-defibrillator (ICD) is highly effective in reducing mortality among patients at risk for fatal arrhythmias, but inappropriate ICD activations are frequent, with potential adverse effects.

METHODS:

We randomly assigned 1500 patients with a primary-prevention indication to receive an ICD with one of three programming configurations. The primary objective was to determine whether programmed high-rate therapy (with a 2.5-second delay before the initiation of therapy at a heart rate of ≥200 beats per minute) or delayed therapy (with a 60-second delay at 170 to 199 beats per minute, a 12-second delay at 200 to 249 beats per minute, and a 2.5-second delay at ≥250 beats per minute) was associated with a decrease in the number of patients with a first occurrence of inappropriate antitachycardia pacing or shocks, as compared with conventional programming (with a 2.5-second delay at 170 to 199 beats per minute and a 1.0-second delay at ≥200 beats per minute).

RESULTS:

During an average follow-up of 1.4 years, high-rate therapy and delayed ICD therapy, as compared with conventional device programming, were associated with reductions in a first occurrence of inappropriate therapy (hazard ratio with high-rate therapy vs. conventional therapy, 0.21; 95% confidence interval [CI], 0.13 to 0.34; P<0.001; hazard ratio with delayed therapy vs. conventional therapy, 0.24; 95% CI, 0.15 to 0.40; P<0.001) and reductions in all-cause mortality (hazard ratio with high-rate therapy vs. conventional therapy, 0.45; 95% CI, 0.24 to 0.85; P=0.01; hazard ratio with delayed therapy vs. conventional therapy, 0.56; 95% CI, 0.30 to 1.02; P=0.06). There were no significant differences in procedure-related adverse events among the three treatment groups.

CONCLUSIONS:

Programming of ICD therapies for tachyarrhythmias of 200 beats per minute or higher or with a prolonged delay in therapy at 170 beats per minute or higher, as compared with conventional programming, was associated with reductions in inappropriate therapy and all-cause mortality during long-term follow-up. (Funded by Boston Scientific; MADIT-RIT ClinicalTrials.gov number, NCT00947310.).
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Taquicardia / Desfibriladores Implantáveis Tipo de estudo: Clinical_trials / Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: N Engl J Med Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Taquicardia / Desfibriladores Implantáveis Tipo de estudo: Clinical_trials / Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: N Engl J Med Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Estados Unidos