Uncertainty evaluation in radon concentration measurement using charcoal canister.
Appl Radiat Isot
; 87: 452-5, 2014 May.
Article
em En
| MEDLINE
| ID: mdl-24444699
ABSTRACT
Active charcoal detectors are used for testing the concentration of radon in dwellings. The method of measurement is based on radon adsorption on coal and measurement of gamma radiation of radon daughters. The contributions to the final measurement uncertainty are identiï¬ed, based on the equation for radon activity concentration calculation. Different methods for setting the region of interest for gamma spectrometry of canisters were discussed and evaluated. The obtained radon activity concentration and uncertainties do not depend on peak area determination method.