[Application of confocal micro-beam X-ray fluorescence in nondestructive scanning analysis of the distribution of elements in a single hair].
Guang Pu Xue Yu Guang Pu Fen Xi
; 33(11): 3147-50, 2013 Nov.
Article
em Zh
| MEDLINE
| ID: mdl-24555400
The confocal micro X-ray fluorescence (XRF) based on polycapillary X-ray lens and conventional X-ray source was used to carry out the scanning analysis of the distribution of the elements in a single hair. The elemental distribution in the single hair was obtained. In the confocal micro XRF technology, the output focal spot of the polycapillary focusing X-ray lens and the input focal spot of the polycapillary parallel X-ray lens were adjusted confocally. The detector could only detect the X-rays from the overlapping foci. This confocal structure decreased the effects of the background on the X-ray spectra, and was accordingly helpful for improving the accuracy of this XRF technology. A polycapillary focusing X-ray lens with a high gain in power density was used to decrease the requirement of power of the X-ray source used in this confocal technology, and made it possible to perform such confocal micro XRF analysis by using the conventional X-ray source with low cost. Experimental results indicated that the confocal micro X-ray fluorescence based on polycapillary X-ray lens had potential applications in analyzing the elemental distribution of individual hairs.