Kuribayashi, S; Iwakiri, K; Kawada, A; Kawami, N; Hoshino, S; Takenouchi, N; Hosaka, H; Shimoyama, Y; Kawamura, O; Yamada, M; Kusano, M.
Afiliação
  • Kuribayashi S; Department of Gastroenterology, Gunma University Hospital, Gunma, Japan; Department of Gastroenterology and Hepatology, National Hospital Organization Numata National Hospital, Gunma, Japan.
Neurogastroenterol Motil ; 27(2): 188-94, 2015 Feb.
Article em En | MEDLINE | ID: mdl-25271562
ABSTRACT

BACKGROUND:

Recently reported normal values for esophageal motility obtained by high-resolution manometry (HRM) using a system with a Unisensor catheter were significantly different from those obtained by the ManoScan(®) , which could result in a wrong diagnosis. To clarify whether these differences were due to system or subject differences, we compared the manometric parameter values between ManoScan and a new system with a Unisensor catheter (Starlet) in the same subjects.

METHODS:

A total of 103 volunteers without any symptoms related to esophageal motility disorders were recruited. Esophageal HRM was performed using both the ManoScan and the Starlet in all subjects. Data from the ManoScan were analyzed using ManoView, and data from the Starlet were analyzed by a program with e-sleeve function. Integrated relaxation pressure, distal contractile integral, contractile front velocity (CFV), intrabolus pressure, and distal latency were calculated by both analyzing programs, and the values of these parameters were compared between the two systems by a signed rank test. KEY

RESULTS:

Data from a total of 97 participants were analyzed. The values of all parameters, except CFV, measured by the Starlet were significantly higher than those obtained by the ManoScan (p < 0.01). CONCLUSIONS & INFERENCES Both systems can measure esophageal motility appropriately; nevertheless, we confirmed that the two systems showed different values of the parameters defined by the Chicago criteria. These differences should be recognized to evaluate esophageal motility precisely.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Transtornos da Motilidade Esofágica / Esôfago / Motilidade Gastrointestinal / Manometria Limite: Humans Idioma: En Revista: Neurogastroenterol Motil Assunto da revista: GASTROENTEROLOGIA / NEUROLOGIA Ano de publicação: 2015 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Transtornos da Motilidade Esofágica / Esôfago / Motilidade Gastrointestinal / Manometria Limite: Humans Idioma: En Revista: Neurogastroenterol Motil Assunto da revista: GASTROENTEROLOGIA / NEUROLOGIA Ano de publicação: 2015 Tipo de documento: Article País de afiliação: Japão