Your browser doesn't support javascript.
loading
Cardiac implantable electronic device remote monitoring surveillance in pediatric and congenital heart disease: Utility relative to frequency.
Dechert, Brynn E; Serwer, Gerald A; Bradley, David J; Dick, Macdonald; LaPage, Martin J.
Afiliação
  • Dechert BE; Department of Pediatrics and Communicable Diseases, Division of Pediatric Cardiology, University of Michigan, Ann Arbor, Michigan. Electronic address: brynnd@umich.edu.
  • Serwer GA; Department of Pediatrics and Communicable Diseases, Division of Pediatric Cardiology, University of Michigan, Ann Arbor, Michigan.
  • Bradley DJ; Department of Pediatrics and Communicable Diseases, Division of Pediatric Cardiology, University of Michigan, Ann Arbor, Michigan.
  • Dick M; Department of Pediatrics and Communicable Diseases, Division of Pediatric Cardiology, University of Michigan, Ann Arbor, Michigan.
  • LaPage MJ; Department of Pediatrics and Communicable Diseases, Division of Pediatric Cardiology, University of Michigan, Ann Arbor, Michigan.
Heart Rhythm ; 12(1): 117-22, 2015 Jan.
Article em En | MEDLINE | ID: mdl-25304681
ABSTRACT

BACKGROUND:

Heart Rhythm Society expert consensus provides recommendations for surveillance monitoring of cardiac implantable electronic devices (CIEDs), but limited data are available for the pediatric and congenital heart disease population.

OBJECTIVE:

The purpose of this study was to determine the rate of actionable events during CIED surveillance in this population, assess the utility of routine CIED remote monitoring, and assess the potential benefit from more frequent monitoring.

METHODS:

This was a retrospective cohort study of all CIED patients followed at a pediatric and congenital heart center and enrolled in the Medtronic Carelink system with either (1) a chronic (implanted for >6 months) CIED followed between July 1, 2010, and July 1, 2012, on a bimonthly schedule; or (2) a new CIED (implanted <6 months) between July 1, 2008, and July 1, 2012, followed on a monthly schedule.

RESULTS:

Aggregate mean age was 20 ± 13.7 years. There were 608 interrogations on newly implanted CIEDs with an actionable event rate of 11 per 100 patient-years. There were 2614 interrogations of chronic CIEDs with an actionable event rate of 22 per 100 patient-years. The odds of an actionable event on an asymptomatic remote monitoring transmission was lower than if symptomatic (odds ratio 0.04, 95% confidence interval 0.03-0.07). Tachyarrhythmia was the most common event. Predictors of actionable events were identified.

CONCLUSION:

In this population of pediatric and congenital heart disease patients, the rate of actionable events was low, especially on asymptomatic interrogations. Most actionable events were due to tachyarrhythmia. A more frequent than every 90-day monitoring schedule does not appear to be of significant benefit.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Arritmias Cardíacas / Desfibriladores Implantáveis / Tecnologia de Sensoriamento Remoto / Cardiopatias Congênitas Tipo de estudo: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies / Screening_studies Limite: Adolescent / Adult / Child / Child, preschool / Female / Humans / Male Idioma: En Revista: Heart Rhythm Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Arritmias Cardíacas / Desfibriladores Implantáveis / Tecnologia de Sensoriamento Remoto / Cardiopatias Congênitas Tipo de estudo: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies / Screening_studies Limite: Adolescent / Adult / Child / Child, preschool / Female / Humans / Male Idioma: En Revista: Heart Rhythm Ano de publicação: 2015 Tipo de documento: Article