Your browser doesn't support javascript.
loading
Experimental study on the mechanical interaction between silicon neural microprobes and rat dura mater during insertion.
Fekete, Z; Németh, A; Márton, G; Ulbert, I; Pongrácz, A.
Afiliação
  • Fekete Z; MEMS Lab, Institute for Technical Physics & Material Science, RCNS, HAS, P.O.Box 49, Budapest, 1525, Hungary, fekete@mfa.kfki.hu.
J Mater Sci Mater Med ; 26(2): 70, 2015 Feb.
Article em En | MEDLINE | ID: mdl-25631267
ABSTRACT
In vivo insertion experiments are essential to optimize novel neural implants. Our work focuses on the interaction between intact dura mater of rats and as-fabricated single-shaft silicon microprobes realized by deep reactive ion etching. Implantation parameters like penetration force and dimpling through intact dura mater were studied as a function of insertion speed, microprobe cross-section, tip angle and animal age. To reduce tissue resistance, we proposed a unique tip sharpening technique, which was also evaluated in in vivo insertion tests. By doubling the insertion speed (between 1.2 and 10.5 mm/min), an increase of 10-35% in penetration forces was measured. When decreasing the cross-section of the microprobes, penetration forces and dimpling was reduced by as much as 30-50% at constant insertion speeds. Force was noticed to gradually decrease by decreasing tip angles. Measured penetration forces through dura mater were reduced even down to 11±3 mN compared to unsharpened (49±13 mN) probes by utilizing our unique tip sharpening technique, which is very close to exerted penetration force in the case of retracted dura (5±1.5 mN). Our findings imply that age remarkably alters the elasticity of intact dura mater. The decreasing stiffness of dura mater results in a significant rise in penetration force and decrease in dimpling. Our work is the first in vivo comparative study on microelectrode penetration through intact and retracted dura mater.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Punção Espinal / Implantação de Prótese / Dura-Máter / Microeletrodos Limite: Animals Idioma: En Revista: J Mater Sci Mater Med Assunto da revista: ENGENHARIA BIOMEDICA Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Punção Espinal / Implantação de Prótese / Dura-Máter / Microeletrodos Limite: Animals Idioma: En Revista: J Mater Sci Mater Med Assunto da revista: ENGENHARIA BIOMEDICA Ano de publicação: 2015 Tipo de documento: Article