Your browser doesn't support javascript.
loading
Quantitative assessment of amide proton transfer (APT) and nuclear overhauser enhancement (NOE) imaging with extrapolated semisolid magnetization transfer reference (EMR) signals: II. Comparison of three EMR models and application to human brain glioma at 3 Tesla.
Heo, Hye-Young; Zhang, Yi; Jiang, Shanshan; Lee, Dong-Hoon; Zhou, Jinyuan.
Afiliação
  • Heo HY; Divison of MR Research, Department of Radiology, Johns Hopkins University, Baltimore, Maryland, USA.
  • Zhang Y; Divison of MR Research, Department of Radiology, Johns Hopkins University, Baltimore, Maryland, USA.
  • Jiang S; Divison of MR Research, Department of Radiology, Johns Hopkins University, Baltimore, Maryland, USA.
  • Lee DH; Divison of MR Research, Department of Radiology, Johns Hopkins University, Baltimore, Maryland, USA.
  • Zhou J; Divison of MR Research, Department of Radiology, Johns Hopkins University, Baltimore, Maryland, USA.
Magn Reson Med ; 75(4): 1630-9, 2016 Apr.
Article em En | MEDLINE | ID: mdl-26033553
ABSTRACT

PURPOSE:

To evaluate the use of three extrapolated semisolid magnetization transfer reference (EMR) methods to quantify amide proton transfer (APT) and nuclear Overhauser enhancement (NOE) signals in human glioma.

METHODS:

Eleven patients with high-grade glioma were scanned at 3 Tesla. aEMR(2) (asymmetric magnetization-transfer or MT model to fit two-sided, wide-offset data), sEMR(2) (symmetric MT model to fit two-sided, wide-offset data), and sEMR(1) (symmetric MT model to fit one-sided, wide-offset data) were assessed. ZEMR and experimental data at 3.5 ppm and -3.5 ppm were subtracted to calculate the APT and NOE signals (APT(#) and NOE(#)), respectively.

RESULTS:

The aEMR(2) and sEMR(1) models provided quite similar APT(#) signals, while the sEMR(2) provided somewhat lower APT(#) signals. The aEMR(2) had an erroneous NOE(#) quantification. Calculated APT(#) signal intensities of glioma (∼4%), much larger than the values reported previously, were significantly higher than those of edema and normal tissue. Compared with normal tissue, gadolinium-enhancing tumor cores were consistently hyperintense on the APT(#) maps and slightly hypointense on the NOE(#) maps.

CONCLUSION:

The sEMR(1) model is the best choice for accurately quantifying APT and NOE signals. The APT-weighted hyperintensity in the tumor was dominated by the APT effect, and the MT asymmetry at 3.5 ppm is a reliable and valid metric for APT imaging of gliomas at 3T.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Neoplasias Encefálicas / Imageamento por Ressonância Magnética / Glioma Tipo de estudo: Prognostic_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Neoplasias Encefálicas / Imageamento por Ressonância Magnética / Glioma Tipo de estudo: Prognostic_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Estados Unidos