Your browser doesn't support javascript.
loading
Better Axial Stiffness of a Bicortical Screw Construct Compared to a Cable Construct for Comminuted Vancouver B1 Proximal Femoral Fractures.
Griffiths, Jamie T; Taheri, Arash; Day, Robert E; Yates, Piers J.
Afiliação
  • Griffiths JT; Fremantle Hospital, Fremantle, Western Australia.
  • Taheri A; Fremantle Hospital, Fremantle, Western Australia.
  • Day RE; Medical Engineering and Physics Department, Royal Perth Hospital, Perth, Western Australia, Australia.
  • Yates PJ; Fremantle Hospital, Fremantle, Western Australia.
J Arthroplasty ; 30(12): 2333-7, 2015 Dec.
Article em En | MEDLINE | ID: mdl-26239233
ABSTRACT
The aim of this study was to biomechanically evaluate the Locking attachment plate (LAP) construct in comparison to a Cable plate construct, for the fixation of periprosthetic femoral fractures after cemented total hip arthroplasty. Each construct incorporated a locking compression plate with bi-cortical locking screws for distal fixation. In the Cable construct, 2 cables and 2 uni-cortical locking screws were used for proximal fixation. In the LAP construct, the cables were replaced by a LAP with 4 bi-cortical locking screws. The LAP construct was significantly stiffer than the cable construct under axial load with a bone gap (P=0.01). The LAP construct offers better axial stiffness compared to the cable construct in the fixation of comminuted Vancouver B1 proximal femoral fractures.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Parafusos Ósseos / Fios Ortopédicos / Fraturas Cominutivas / Artroplastia de Quadril / Fraturas do Fêmur / Fixação Interna de Fraturas Tipo de estudo: Clinical_trials / Etiology_studies / Evaluation_studies Limite: Humans Idioma: En Revista: J Arthroplasty Assunto da revista: ORTOPEDIA Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Parafusos Ósseos / Fios Ortopédicos / Fraturas Cominutivas / Artroplastia de Quadril / Fraturas do Fêmur / Fixação Interna de Fraturas Tipo de estudo: Clinical_trials / Etiology_studies / Evaluation_studies Limite: Humans Idioma: En Revista: J Arthroplasty Assunto da revista: ORTOPEDIA Ano de publicação: 2015 Tipo de documento: Article