Determination of optimal imaging settings for urolithiasis CT using filtered back projection (FBP), statistical iterative reconstruction (IR) and knowledge-based iterative model reconstruction (IMR): a physical human phantom study.
Br J Radiol
; 89(1058): 20150527, 2016.
Article
em En
| MEDLINE
| ID: mdl-26577542
ABSTRACT
OBJECTIVE:
The purpose of this study was to compare CT image quality for evaluating urolithiasis using filtered back projection (FBP), statistical iterative reconstruction (IR) and knowledge-based iterative model reconstruction (IMR) according to various scan parameters and radiation doses.METHODS:
A 5 × 5 × 5 mm(3) uric acid stone was placed in a physical human phantom at the level of the pelvis. 3 tube voltages (120, 100 and 80 kV) and 4 current-time products (100, 70, 30 and 15 mAs) were implemented in 12 scans. Each scan was reconstructed with FBP, statistical IR (Levels 5-7) and knowledge-based IMR (soft-tissue Levels 1-3). The radiation dose, objective image quality and signal-to-noise ratio (SNR) were evaluated, and subjective assessments were performed.RESULTS:
The effective doses ranged from 0.095 to 2.621 mSv. Knowledge-based IMR showed better objective image noise and SNR than did FBP and statistical IR. The subjective image noise of FBP was worse than that of statistical IR and knowledge-based IMR. The subjective assessment scores deteriorated after a break point of 100 kV and 30 mAs.CONCLUSION:
At the setting of 100 kV and 30 mAs, the radiation dose can be decreased by approximately 84% while keeping the subjective image assessment. ADVANCES IN KNOWLEDGE Patients with urolithiasis can be evaluated with ultralow-dose non-enhanced CT using a knowledge-based IMR algorithm at a substantially reduced radiation dose with the imaging quality preserved, thereby minimizing the risks of radiation exposure while providing clinically relevant diagnostic benefits for patients.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Urografia
/
Interpretação de Imagem Radiográfica Assistida por Computador
/
Tomografia Computadorizada por Raios X
/
Urolitíase
Tipo de estudo:
Prognostic_studies
Limite:
Humans
Idioma:
En
Revista:
Br J Radiol
Ano de publicação:
2016
Tipo de documento:
Article