Your browser doesn't support javascript.
loading
A shear gradient-activated microfluidic device for automated monitoring of whole blood haemostasis and platelet function.
Jain, Abhishek; Graveline, Amanda; Waterhouse, Anna; Vernet, Andyna; Flaumenhaft, Robert; Ingber, Donald E.
Afiliação
  • Jain A; Wyss Institute for Biologically Inspired Engineering, Harvard University, Boston, Massachusetts 02115, USA.
  • Graveline A; Division of Hemostasis and Thrombosis, Department of Medicine, Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, Massachusetts 02115, USA.
  • Waterhouse A; Vascular Biology Program, Departments of Pathology and Surgery, Boston Children's Hospital and Harvard Medical School, Boston, Massachusetts 02115, USA.
  • Vernet A; Wyss Institute for Biologically Inspired Engineering, Harvard University, Boston, Massachusetts 02115, USA.
  • Flaumenhaft R; Wyss Institute for Biologically Inspired Engineering, Harvard University, Boston, Massachusetts 02115, USA.
  • Ingber DE; Wyss Institute for Biologically Inspired Engineering, Harvard University, Boston, Massachusetts 02115, USA.
Nat Commun ; 7: 10176, 2016 Jan 06.
Article em En | MEDLINE | ID: mdl-26733371
ABSTRACT
Accurate assessment of blood haemostasis is essential for the management of patients who use extracorporeal devices, receive anticoagulation therapy or experience coagulopathies. However, current monitoring devices do not measure effects of haemodynamic forces that contribute significantly to platelet function and thrombus formation. Here we describe a microfluidic device that mimics a network of stenosed arteriolar vessels, permitting evaluation of blood clotting within small sample volumes under pathophysiological flow. By applying a clotting time analysis based on a phenomenological mathematical model of thrombus formation, coagulation and platelet function can be accurately measured in vitro in patient blood samples. When the device is integrated into an extracorporeal circuit in pig endotoxemia or heparin therapy models, it produces real-time readouts of alterations in coagulation ex vivo that are more reliable than standard clotting assays. Thus, this disposable device may be useful for personalized diagnostics and for real-time surveillance of antithrombotic therapy in clinic.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Testes de Coagulação Sanguínea / Plaquetas / Dispositivos Lab-On-A-Chip / Hemostasia Tipo de estudo: Qualitative_research Limite: Animals / Female / Humans Idioma: En Revista: Nat Commun Assunto da revista: BIOLOGIA / CIENCIA Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Testes de Coagulação Sanguínea / Plaquetas / Dispositivos Lab-On-A-Chip / Hemostasia Tipo de estudo: Qualitative_research Limite: Animals / Female / Humans Idioma: En Revista: Nat Commun Assunto da revista: BIOLOGIA / CIENCIA Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Estados Unidos