Implantable Cardioverter Defibrillator Outcomes in Pediatric and Congenital Heart Disease: Time to System Revision.
Dechert, Brynn E; Bradley, David J; Serwer, Gerald A; Dick Ii, Macdonald; Lapage, Martin J.
Afiliação
  • Dechert BE; Department of Pediatrics and Communicable Diseases, Division of Pediatric Cardiology, University of Michigan, Ann Arbor, Michigan.
  • Bradley DJ; Department of Pediatrics and Communicable Diseases, Division of Pediatric Cardiology, University of Michigan, Ann Arbor, Michigan.
  • Serwer GA; Department of Pediatrics and Communicable Diseases, Division of Pediatric Cardiology, University of Michigan, Ann Arbor, Michigan.
  • Dick Ii M; Department of Pediatrics and Communicable Diseases, Division of Pediatric Cardiology, University of Michigan, Ann Arbor, Michigan.
  • Lapage MJ; Department of Pediatrics and Communicable Diseases, Division of Pediatric Cardiology, University of Michigan, Ann Arbor, Michigan.
Pacing Clin Electrophysiol ; 39(7): 703-8, 2016 Jul.
Article em En | MEDLINE | ID: mdl-27119790
ABSTRACT

BACKGROUND:

Implantable cardioverter defibrillators (ICDs) are intended to prevent sudden cardiac death yet also impose a risk of morbidity. This study describes the outcomes of ICDs in a pediatric and congenital heart disease (CHD) population from a single center.

METHODS:

Retrospective cohort study of all patients with an ICD followed at the University of Michigan Congenital Heart Center from 2005-2013. The primary outcome was ICD system revision for any reason excluding routine generator change for battery depletion.

RESULTS:

There were 191 ICD systems in 131 patients, including 57 with CHD, 24 with hypertrophic cardiomyopathy, and 45 with structurally normal hearts. Median age was 16 years at initial implant. Total follow-up was 850 patient-years; median 4.9 years/patient. There were 43 (33%) patients who required 60 ICD revisions; 70 revisions/1,000 patient-years of follow-up. Revisions included 25 lead extractions with replacement, 21 lead additions, five lead repositions, and four full system revisions. Kaplan-Meier (K-M) median time to appropriate shock was similar to the median time to system revision. K-M time to system revision was significantly affected by recalled lead performance.

CONCLUSIONS:

The need for ICD system revision is high in this pediatric and CHD population and occurs at a rate similar to the rate of receiving appropriate therapy. These results highlight the need for judicious implant criteria and improved device longevity.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Desfibriladores Implantáveis / Remoção de Dispositivo / Cardiopatias Congênitas / Insuficiência Cardíaca Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Prevalence_studies / Risk_factors_studies Limite: Child / Child, preschool / Female / Humans / Male País/Região como assunto: America do norte Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Desfibriladores Implantáveis / Remoção de Dispositivo / Cardiopatias Congênitas / Insuficiência Cardíaca Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Prevalence_studies / Risk_factors_studies Limite: Child / Child, preschool / Female / Humans / Male País/Região como assunto: America do norte Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2016 Tipo de documento: Article