Improved standardization and potential for shortened time to results with BD Kiestra™ total laboratory automation of early urine cultures: A prospective comparison with manual processing.
Diagn Microbiol Infect Dis
; 86(1): 1-4, 2016 Sep.
Article
em En
| MEDLINE
| ID: mdl-27422083
ABSTRACT
We compared the results of 505 urine specimens prospectively processed by both conventional manual processing (MP) with 16-24h incubation to BD Kiestra™ Total Laboratory Automation (TLA) system with a shortened incubation of 14h 97% of culture results were clinically concordant. TLA processing was associated with improved standardization of time of first culture reading and total incubation time.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Infecções Urinárias
/
Técnicas Bacteriológicas
/
Automação Laboratorial
Tipo de estudo:
Evaluation_studies
/
Guideline
/
Observational_studies
Limite:
Humans
Idioma:
En
Revista:
Diagn Microbiol Infect Dis
Ano de publicação:
2016
Tipo de documento:
Article