ToF-SIMS and Principal Component Analysis Investigation of Denatured, Surface-Adsorbed Antibodies.
Welch, Nicholas G; Madiona, Robert M T; Scoble, Judith A; Muir, Benjamin W; Pigram, Paul J.
Afiliação
  • Welch NG; Centre for Materials and Surface Science and Department of Chemistry and Physics, School of Molecular Sciences, La Trobe University , Melbourne, VIC 3086, Australia.
  • Madiona RMT; CSIRO Manufacturing , Clayton, VIC 3168, Australia.
  • Scoble JA; Centre for Materials and Surface Science and Department of Chemistry and Physics, School of Molecular Sciences, La Trobe University , Melbourne, VIC 3086, Australia.
  • Muir BW; CSIRO Manufacturing , Clayton, VIC 3168, Australia.
  • Pigram PJ; CSIRO Manufacturing , Clayton, VIC 3168, Australia.
Langmuir ; 32(42): 10824-10834, 2016 Oct 25.
Article em En | MEDLINE | ID: mdl-27715065
ABSTRACT
Antibody denaturation at solid-liquid interfaces plays an important role in the sensitivity of protein assays such as enzyme-linked immunosorbent assays (ELISAs). Surface immobilized antibodies must maintain their native state, with their antigen binding (Fab) region intact, to capture antigens from biological samples and permit disease detection. In this work, two identical sample sets were prepared with whole antibody IgG, F(ab')2 and Fc fragments, immobilized to either a silicon wafer or a diethylene glycol dimethyl ether plasma polymer surface. Analysis was conducted on one sample set at day 0, and the second sample set after 14 days in vacuum, with time-of-flight secondary ion mass spectrometry (ToF-SIMS) for molecular species representative of denaturation. A 1003 mass fragment peak list was compiled from ToF-SIMS data and compared to a 35 amino acid mass fragment peak list using principal component analysis. Several ToF-SIMS secondary ions, pertaining to disulfide and thiol species, were identified in the 14 day (presumably denatured) samples. A substrate and primary ion independent marker for denaturation (aging) was then produced using a ratio of mass peak intensities according to denaturation ratio [I61.9534 + I62.9846 + I122.9547 + I84.9609 + I120.9461]/[I30.9979 + I42.9991 + I73.0660 + I147.0780]. The ratio successfully identifies denaturation on both the silicon and plasma polymer substrates and for spectra generated with Mn+, Bi+, and Bi3+ primary ions. We believe this ratio could be employed to as a marker of denaturation of antibodies on a plethora of substrates.

Texto completo: 1 Base de dados: MEDLINE Idioma: En Revista: Langmuir Assunto da revista: QUIMICA Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Austrália