Your browser doesn't support javascript.
loading
Maintaining Accurate Long-Term Sensing Ability Despite Significant Size Reduction of Implantable Cardiac Monitors.
Lortz, Julia; Varnavas, Varnavas; WEIßENBERGER, Wilko; Erbel, Raimund; Reinsch, Nico.
Afiliação
  • Lortz J; Department of Cardiology, West-German Heart and Vascular Center, University Hospital Essen, Essen, Germany.
  • Varnavas V; Department of Cardiology, West-German Heart and Vascular Center, University Hospital Essen, Essen, Germany.
  • WEIßENBERGER W; Department of Thoracic and Cardiovascular Surgery, West-German Heart and Vascular Center, University Hospital Essen, Essen, Germany.
  • Erbel R; Institute of Medical Informatics, Biometry and Epidemiology, University Hospital Essen, Essen, Germany.
  • Reinsch N; Division of Electrophysiology, Department of Internal Medicine I and Cardiology, Alfried Krupp Hospital Essen, Essen, Germany.
Pacing Clin Electrophysiol ; 39(12): 1344-1350, 2016 Dec.
Article em En | MEDLINE | ID: mdl-27862030
ABSTRACT

BACKGROUND:

The Reveal LINQ™ implantable cardiac monitor (ICM; Medtronic, Minneapolis, MN, USA) is obviously smaller than its precursor (the Reveal XT™), but little is known about its long-term safety. Here, we investigated the long-term R-wave sensing reliability of the Reveal LINQ™ ICM.

METHODS:

We analyzed the sensing quality of the Reveal LINQ™ ICM over time between March 2014 and January 2015.

RESULTS:

A total of 30 patients underwent Reveal LINQ™ implantation. The main reason for implantation was unexplained syncope (60%). We evaluated a total of 305.8 patient-months (median, 10.7 months) of R-wave sensing quality. A comparison of baseline and follow-up R-wave sensing values revealed no difference (0.401 mV ± 0.244 mV vs 0.447 mV ± 0.323 mV; P = 0.225).

CONCLUSIONS:

The consistent detection of events is an important safety feature of an ICM and linked to secure R-wave sensing. Despite the noticeable size reduction, the Reveal LINQ™ ICM retains reliable sensing quality over time.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Arritmias Cardíacas / Próteses e Implantes / Eletrocardiografia Ambulatorial / Satisfação do Paciente / Segurança de Equipamentos / Tecnologia sem Fio Tipo de estudo: Diagnostic_studies / Observational_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Arritmias Cardíacas / Próteses e Implantes / Eletrocardiografia Ambulatorial / Satisfação do Paciente / Segurança de Equipamentos / Tecnologia sem Fio Tipo de estudo: Diagnostic_studies / Observational_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Alemanha