Your browser doesn't support javascript.
loading
Spatiotemporal alignment of in utero BOLD-MRI series.
Turk, Esra Abaci; Luo, Jie; Gagoski, Borjan; Pascau, Javier; Bibbo, Carolina; Robinson, Julian N; Grant, P Ellen; Adalsteinsson, Elfar; Golland, Polina; Malpica, Norberto.
Afiliação
  • Turk EA; Fetal-Neonatal Neuroimaging and Developmental Science Center, Boston Children's Hospital, Harvard Medical School, Boston, MA, United States.
  • Luo J; Madrid-MIT M+Vision Consortium in RLE, Massachusetts Institute of Technology, Cambridge, MA, United States.
  • Gagoski B; Fetal-Neonatal Neuroimaging and Developmental Science Center, Boston Children's Hospital, Harvard Medical School, Boston, MA, United States.
  • Pascau J; Madrid-MIT M+Vision Consortium in RLE, Massachusetts Institute of Technology, Cambridge, MA, United States.
  • Bibbo C; Fetal-Neonatal Neuroimaging and Developmental Science Center, Boston Children's Hospital, Harvard Medical School, Boston, MA, United States.
  • Robinson JN; Department of Radiology, Harvard Medical School, Boston Children's Hospital, Boston, MA, United States.
  • Grant PE; Madrid-MIT M+Vision Consortium in RLE, Massachusetts Institute of Technology, Cambridge, MA, United States.
  • Adalsteinsson E; Instituto de Investigación Sanitaria Gregorio Marañón, CIBERSAM, Madrid, Spain.
  • Golland P; Departamento de Bioingeniería e Ingeniería Aeroespacial, Universidad Carlos III de Madrid, Madrid, Spain.
  • Malpica N; Maternal and Fetal Medicine, Brigham and Women's Hospital, Harvard Medical School, Boston, MA, United States.
J Magn Reson Imaging ; 46(2): 403-412, 2017 08.
Article em En | MEDLINE | ID: mdl-28152240
ABSTRACT

PURPOSE:

To present a method for spatiotemporal alignment of in-utero magnetic resonance imaging (MRI) time series acquired during maternal hyperoxia for enabling improved quantitative tracking of blood oxygen level-dependent (BOLD) signal changes that characterize oxygen transport through the placenta to fetal organs. MATERIALS AND

METHODS:

The proposed pipeline for spatiotemporal alignment of images acquired with a single-shot gradient echo echo-planar imaging includes 1) signal nonuniformity correction, 2) intravolume motion correction based on nonrigid registration, 3) correction of motion and nonrigid deformations across volumes, and 4) detection of the outlier volumes to be discarded from subsequent analysis. BOLD MRI time series collected from 10 pregnant women during 3T scans were analyzed using this pipeline. To assess pipeline performance, signal fluctuations between consecutive timepoints were examined. In addition, volume overlap and distance between manual region of interest (ROI) delineations in a subset of frames and the delineations obtained through propagation of the ROIs from the reference frame were used to quantify alignment accuracy. A previously demonstrated rigid registration approach was used for comparison.

RESULTS:

The proposed pipeline improved anatomical alignment of placenta and fetal organs over the state-of-the-art rigid motion correction methods. In particular, unexpected temporal signal fluctuations during the first normoxia period were significantly decreased (P < 0.01) and volume overlap and distance between region boundaries measures were significantly improved (P < 0.01).

CONCLUSION:

The proposed approach to align MRI time series enables more accurate quantitative studies of placental function by improving spatiotemporal alignment across placenta and fetal organs. LEVEL OF EVIDENCE 1 Technical Efficacy Stage 1 J. MAGN. RESON. IMAGING 2017;46403-412.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Oxigênio / Placenta / Processamento de Imagem Assistida por Computador / Imageamento por Ressonância Magnética / Feto Limite: Female / Humans / Pregnancy Idioma: En Revista: J Magn Reson Imaging Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Oxigênio / Placenta / Processamento de Imagem Assistida por Computador / Imageamento por Ressonância Magnética / Feto Limite: Female / Humans / Pregnancy Idioma: En Revista: J Magn Reson Imaging Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Estados Unidos