Biomechanical comparison of anatomical plating systems for comminuted distal humeral fractures.
Varady, Patrick A; von Rüden, Christian; Greinwald, Markus; Hungerer, Sven; Pätzold, Robert; Augat, Peter.
Afiliação
  • Varady PA; Institute of Biomechanics, Paracelsus Medical University, Salzburg, Austria. p.varady@ibmb.tu-bs.de.
  • von Rüden C; Institute of Biomechanics, Trauma Centre Murnau, Professor Küntscher Str. 8, 82418, Murnau, Germany. p.varady@ibmb.tu-bs.de.
  • Greinwald M; Institute of Biomechanics, Paracelsus Medical University, Salzburg, Austria.
  • Hungerer S; Institute of Biomechanics, Trauma Centre Murnau, Professor Küntscher Str. 8, 82418, Murnau, Germany.
  • Pätzold R; Department of Trauma Surgery, Trauma Centre Murnau, Professor Küntscher Str. 8, 82418, Murnau, Germany.
  • Augat P; Institute of Biomechanics, Paracelsus Medical University, Salzburg, Austria.
Int Orthop ; 41(9): 1709-1714, 2017 09.
Article em En | MEDLINE | ID: mdl-28349182
ABSTRACT

PURPOSE:

Six different mono-axial and poly-axial distal humeral plating systems with an anatomical plate design were compared. The aim of the biomechanical tests was to examine differences regarding system stiffness, median fatigue limit, and failure mechanisms.

METHODS:

Different configurations of two double plate fixation systems by two manufacturers for the treatment of complex distal humeral fractures (AO/OTA type C2.3) were biomechanically tested in a physiologically relevant setup.

RESULTS:

The 180° Stryker configuration presented itself as the system with the highest stiffness, being significantly stiffer (p < 0.001) than every system other than the poly-axial 180° aap system (p = 0.378). For the median fatigue limit the 180° Stryker and poly-axial aap systems were ranked first and second. The failure mechanism for all 90° systems was a fatigue breakage of the posterolateral plate. The 180° aap systems demonstrated breakage of the most distal screws of the lateral plate. The 180° Stryker system demonstrated screw breakage on both the medial and lateral plates.

DISCUSSION:

Breakage of the posterolateral plate as a failure mechanism for the 90° systems was expected. The 180° systems demonstrated a higher stiffness compared to the 90° constructs for the axial loading. In conclusion, both poly-axial anatomical plating systems provide sufficient stability in this scenario, and the 180° configurations demonstrated superior stiffness.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Placas Ósseas / Fraturas Cominutivas / Fixação Interna de Fraturas / Fraturas do Úmero / Úmero Tipo de estudo: Etiology_studies Limite: Humans Idioma: En Revista: Int Orthop Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Áustria

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Placas Ósseas / Fraturas Cominutivas / Fixação Interna de Fraturas / Fraturas do Úmero / Úmero Tipo de estudo: Etiology_studies Limite: Humans Idioma: En Revista: Int Orthop Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Áustria