Your browser doesn't support javascript.
loading
A Novel Noncontact Diffuse Correlation Spectroscopy Device for Assessing Blood Flow in Mastectomy Skin Flaps: A Prospective Study in Patients Undergoing Prosthesis-Based Reconstruction.
Agochukwu, Nneamaka B; Huang, Chong; Zhao, Mingjun; Bahrani, Ahmed A; Chen, Li; McGrath, Patrick; Yu, Guoqiang; Wong, Lesley.
Afiliação
  • Agochukwu NB; Lexington, Ky.
  • Huang C; From the Division of Plastic Surgery, the Department of Biomedical Engineering, the Department of Biostatistics, the Markey Cancer Center, Biostatistics Shared Resource Facility, and the Division of General Surgery, University of Kentucky.
  • Zhao M; Lexington, Ky.
  • Bahrani AA; From the Division of Plastic Surgery, the Department of Biomedical Engineering, the Department of Biostatistics, the Markey Cancer Center, Biostatistics Shared Resource Facility, and the Division of General Surgery, University of Kentucky.
  • Chen L; Lexington, Ky.
  • McGrath P; From the Division of Plastic Surgery, the Department of Biomedical Engineering, the Department of Biostatistics, the Markey Cancer Center, Biostatistics Shared Resource Facility, and the Division of General Surgery, University of Kentucky.
  • Yu G; Lexington, Ky.
  • Wong L; From the Division of Plastic Surgery, the Department of Biomedical Engineering, the Department of Biostatistics, the Markey Cancer Center, Biostatistics Shared Resource Facility, and the Division of General Surgery, University of Kentucky.
Plast Reconstr Surg ; 140(1): 26-31, 2017 Jul.
Article em En | MEDLINE | ID: mdl-28654584
ABSTRACT
A new advanced technology, noncontact diffuse correlation spectroscopy, has been recently developed for the measurement of tissue blood flow through analyzing the motions of red blood cells in deep tissues. This technology is portable, inexpensive, and noninvasive, and can measure up to 1.5-cm tissue depth. In this prospective study, the authors aimed to explore the use of this novel device in the prediction of mastectomy skin flap necrosis. The noncontact diffuse correlation spectroscopy device was used to measure mastectomy skin flap flow in patients undergoing mastectomy and immediate implant-based breast reconstruction before and immediately after mastectomy, and after placement of the prosthesis. Patients were tracked for the development of complications, including skin necrosis and the need for further surgery. Nineteen patients were enrolled in the study. Four patients (21 percent) developed skin necrosis, one of which required additional surgery. The difference in relative blood flow levels immediately after mastectomy in patients with or without necrosis was statistically significant, with values of 0.27 ± 0.11 and 0.66 ± 0.22, respectively (p = 0.0005). Relative blood flow measurements immediately after mastectomy show a significant high accuracy in prediction of skin flap necrosis, with an area under the receiver operating characteristic curve of 0.95 (95 percent confidence interval, 0.81 to 1). The noncontact diffuse correlation spectroscopy device is a promising tool that provides objective information regarding mastectomy skin flap viability intraoperatively, allowing surgeons early identification of those compromised and ischemic flaps with the hope of potentially salvaging them. CLINICAL QUESTION/LEVEL OF EVIDENCE Diagnostic, IV.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fluxo Sanguíneo Regional / Análise Espectral / Retalhos Cirúrgicos / Implante Mamário / Mastectomia Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies Limite: Adult / Aged / Female / Humans / Middle aged Idioma: En Revista: Plast Reconstr Surg Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fluxo Sanguíneo Regional / Análise Espectral / Retalhos Cirúrgicos / Implante Mamário / Mastectomia Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies Limite: Adult / Aged / Female / Humans / Middle aged Idioma: En Revista: Plast Reconstr Surg Ano de publicação: 2017 Tipo de documento: Article