A concordance study of the Altona RealStar Varicella-Zoster virus real-time quantitative PCR and in-house conventional qualitative PCR.
J Med Virol
; 90(2): 334-337, 2018 02.
Article
em En
| MEDLINE
| ID: mdl-28971486
This study compared the Altona RealStar™ VZV Kit 1.0 real time quantitative VZV PCR with in-house qualitative conventional VZV PCR on cerebrospinal fluid, mucocutaneous, and other uncommon clinical specimens. Overall, positive and negative agreement percentages were respectively 97.9% (95%CI: 93.8-99.6), 100.0% (95%CI: 93.1-100.0), and 96.3% (95%CI: 89.4-99.2) while Cohen's kappa statistic value was 0.96 (95%CI: 0.91-1.00). RealStar™ VZV quantitative PCR assay reported average quantitative viral loads of 4.4 × 105 and 1.1 × 107 copies/mL in cerebrospinal fluid and cutaneous specimens, respectively (P < 0.01). RealStar™ VZV PCR assay showed excellent agreement with in house conventional assay for various clinical specimens.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Reação em Cadeia da Polimerase
/
Herpesvirus Humano 3
/
Técnicas de Diagnóstico Molecular
/
Infecção pelo Vírus da Varicela-Zoster
Tipo de estudo:
Diagnostic_studies
/
Evaluation_studies
/
Observational_studies
/
Qualitative_research
Limite:
Humans
Idioma:
En
Revista:
J Med Virol
Ano de publicação:
2018
Tipo de documento:
Article
País de afiliação:
Canadá