A comparative surface topographical analysis of explanted total knee replacement prostheses: Oxidised zirconium vs cobalt chromium femoral components.
Kennard, Emma; Scholes, Susan C; Sidaginamale, Raghavendra; Gangadharan, Rajkumar; Weir, David J; Holland, James; Deehan, David; Joyce, Thomas J.
Afiliação
  • Kennard E; Bioengineering Group, School of Engineering, Newcastle University, Stephenson Building, NE1 7RU Newcastle upon Tyne, England, United Kingdom. Electronic address: e.kennard@newcastle.ac.uk.
  • Scholes SC; Bioengineering Group, School of Engineering, Newcastle University, Stephenson Building, NE1 7RU Newcastle upon Tyne, England, United Kingdom.
  • Sidaginamale R; Bioengineering Group, School of Engineering, Newcastle University, Stephenson Building, NE1 7RU Newcastle upon Tyne, England, United Kingdom.
  • Gangadharan R; Freeman Hospital, NE1 7DN Newcastle upon Tyne, England, United Kingdom.
  • Weir DJ; Freeman Hospital, NE1 7DN Newcastle upon Tyne, England, United Kingdom.
  • Holland J; Freeman Hospital, NE1 7DN Newcastle upon Tyne, England, United Kingdom.
  • Deehan D; Freeman Hospital, NE1 7DN Newcastle upon Tyne, England, United Kingdom.
  • Joyce TJ; Bioengineering Group, School of Engineering, Newcastle University, Stephenson Building, NE1 7RU Newcastle upon Tyne, England, United Kingdom.
Med Eng Phys ; 50: 59-64, 2017 12.
Article em En | MEDLINE | ID: mdl-29042144
ABSTRACT
It has been proposed that an increased surface roughness of the femoral components of Total Knee Replacements (TKRs) may be a contributing factor to the accelerated wear of the polyethylene (PE) bearing and ultimately prosthesis failure. Oxidised Zirconium was introduced to the orthopaedic market in an attempt to reduce PE wear associated failures and increase the longevity of the prosthesis. In this study, non-contacting profilometry was used to measure the surface roughness of the femoral components of 6 retrieved TKRs (3 Oxidised Zirconium (OxZr) and 3 Cobalt Chromium alloy (CoCr) femoral components) and 2 as-manufactured femoral components (1 OxZr and 1 CoCr). A semi-quantitative method was used to analyse the damage on the retrieved PE components. The Sa values for the retrieved OxZr femoral components (Sa = 0.093 µm ± 0.014) and for the retrieved CoCr femoral components (Sa = 0.065 µm ± 0.005) were significantly greater (p < .05) than the roughness values for the as-manufactured femoral components (OxZr Sa = 0.061 µm ± 0.004 and CoCr Sa = 0.042 µm ± 0.003). No significant difference was seen between the surface roughness parameters of the retrieved OxZr and retrieved CoCr femoral components. There was no difference between the PE component damage scores for the retrieved OxZr TKRs compared to the retrieved CoCr TKRs. These results agree with other studies that both OxZr and CoCr femoral components roughen during time in vivo but the lack of difference between the surface roughness measurements of the two materials is in contrast to previous topographical reports. Further analysis of retrieved OxZr TKRs is recommended so that a fuller appreciation of their benefits and limitations be obtained.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Zircônio / Teste de Materiais / Ligas de Cromo / Cobalto / Artroplastia do Joelho / Fêmur / Prótese do Joelho Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Med Eng Phys Assunto da revista: BIOFISICA / ENGENHARIA BIOMEDICA Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Zircônio / Teste de Materiais / Ligas de Cromo / Cobalto / Artroplastia do Joelho / Fêmur / Prótese do Joelho Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Med Eng Phys Assunto da revista: BIOFISICA / ENGENHARIA BIOMEDICA Ano de publicação: 2017 Tipo de documento: Article