Do screws and screw holes affect osteolysis in cementless cups using highly crosslinked polyethylene? A 7 to 10-year follow-up case-control study.
Taniguchi, N; Jinno, T; Takada, R; Koga, D; Ando, T; Okawa, A; Haro, H.
Afiliação
  • Taniguchi N; Department of Orthopaedic Surgery, University of Yamanashi, 1110 Shimokato, Chuo, Yamanashi 409-3898, Japan; Department of Orthopaedic Surgery, Tokyo Medical and Dental University, 1-5-45 Yushima, Bunkyo-ku, Tokyo 113-8519, Japan.
  • Jinno T; Department of Orthopaedic Surgery, Tokyo Medical and Dental University, 1-5-45 Yushima, Bunkyo-ku, Tokyo 113-8519, Japan; Department of Rehabilitation Medicine, Tokyo Medical and Dental University, 1-5-45 Yushima, Bunkyo-ku, Tokyo 113-8519, Japan. Electronic address: jinno.orth@tmd.ac.jp.
  • Takada R; Department of Orthopaedic Surgery, Tokyo Medical and Dental University, 1-5-45 Yushima, Bunkyo-ku, Tokyo 113-8519, Japan.
  • Koga D; Department of Orthopaedic Surgery, Tokyo Medical and Dental University, 1-5-45 Yushima, Bunkyo-ku, Tokyo 113-8519, Japan.
  • Ando T; Department of Orthopaedic Surgery, University of Yamanashi, 1110 Shimokato, Chuo, Yamanashi 409-3898, Japan.
  • Okawa A; Department of Orthopaedic Surgery, Tokyo Medical and Dental University, 1-5-45 Yushima, Bunkyo-ku, Tokyo 113-8519, Japan.
  • Haro H; Department of Orthopaedic Surgery, University of Yamanashi, 1110 Shimokato, Chuo, Yamanashi 409-3898, Japan.
Orthop Traumatol Surg Res ; 104(3): 307-315, 2018 05.
Article em En | MEDLINE | ID: mdl-29355741
BACKGROUND: The use of screws and the presence of screw holes may cause acetabular osteolysis and implant loosening in cementless total hip arthroplasty (THA) using conventional polyethylene. In contrast, this issue is not fully understood using highly crosslinked polyethylene (HXLPE), particularly in large comparative study. Therefore, we performed a case-control study to assess the influence of screw usage and screw holes on: (1) implant fixation and osteolysis and (2) polyethylene steady-state wear rate, using cases with HXLPE liners followed up for 7-10 years postoperatively. HYPOTHESIS: The screw usage and screw holes adversely affect the implant fixation and incidence of wear-related osteolysis in THA with HXLPE. PATIENTS AND METHODS: We reviewed 209 primary cementless THAs performed with 26-mm cobalt-chromium heads on HXLPE liners. To compare the effects of the use of screws and the presence of screw holes, the following groups were established: (1) with-screw (n=140); (2) without-screw (n=69); (3) no-hole (n=27) and (4) group in which a cup with screw holes, but no screw was used (n=42). Two adjunct groups (no-hole cups excluded) were established to compare the differences in the two types of HXLPE: (5) remelted group (n=100) and (6) annealed group (n=82). Implant stability and osteolysis were evaluated by plain radiography and computed tomography. The wear rate from 1 year to the final evaluation was measured using plain X-rays and PolyWare Digital software. RESULTS: All cups and stems achieved bony fixation. On CT-scan, no acetabular osteolysis was found, but there were 3 cases with a small area of femoral osteolysis. The mean steady-state wear rate of each group was (1) 0.031±0.022, (2) 0.033±0.035, (3) 0.031±0.024, (4) 0.029±0.018, (5) 0.030±0.018 and (6) 0.034±0.023mm/year, respectively. A comparison of the effects of screw usage or screw holes found no significant between-group differences in the implant stability, prevalence of osteolysis [no acetabular osteolysis and 3/209 at femoral side (1.4%)] and steady-state wear rate. DISCUSSION: This study suggests that there are no adverse effects on the results of THA with HXLPE from the use of cups with screw holes and the use of screws for cup fixation. LEVEL OF EVIDENCE: Level III retrospective case-control study.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Osteólise / Parafusos Ósseos / Falha de Prótese / Artroplastia de Quadril / Polietileno / Prótese de Quadril Tipo de estudo: Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Orthop Traumatol Surg Res Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Osteólise / Parafusos Ósseos / Falha de Prótese / Artroplastia de Quadril / Polietileno / Prótese de Quadril Tipo de estudo: Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Orthop Traumatol Surg Res Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Japão