Biomechanical efficacy of AP, PA lag screws and posterior plating for fixation of posterior malleolar fractures: a three dimensional finite element study.
Anwar, Adeel; Zhang, Zhen; Lv, Decheng; Lv, Gang; Zhao, Zhi; Wang, Yanfeng; Cai, Yue; Qasim, Wasim; Nazir, Muhammad Umar; Lu, Ming.
Afiliação
  • Anwar A; Department of Orthopaedic Surgery, The First Affiliated Hospital of Dalian Medical University, 222 Zhongshan road, 116011, Dalian, Liaoning, People's Republic of China.
  • Zhang Z; Department of Orthopaedic Surgery, The First Affiliated Hospital of Dalian Medical University, 222 Zhongshan road, 116011, Dalian, Liaoning, People's Republic of China. zhangzhendmu@163.com.
  • Lv D; Department of Orthopaedic Surgery, The First Affiliated Hospital of Dalian Medical University, 222 Zhongshan road, 116011, Dalian, Liaoning, People's Republic of China. lvdecheng_dmu@163.com.
  • Lv G; Department of Orthopaedic Surgery, The First Affiliated Hospital of China Medical University, 155 Nanjing north street, 110001, Shenyang, Liaoning, People's Republic of China.
  • Zhao Z; Department of Orthopaedic Surgery, The Second Affiliated Hospital of Dalian Medical University, 456 Zhongshan road, 116027, Dalian, Liaoning, People's Republic of China.
  • Wang Y; Department of Orthopaedic Surgery, The First Affiliated Hospital of China Medical University, 155 Nanjing north street, 110001, Shenyang, Liaoning, People's Republic of China.
  • Cai Y; Department of Automation, School of Electrical Engineering, Dalian Jiaotong University, 794 Huanghe road, 116028, Dalian, Liaoning, People's Republic of China.
  • Qasim W; Department of General Surgery, The Second Affiliated Hospital of Dalian Medical University, 456 Zhongshan road, 116027, Dalian, Liaoning, People's Republic of China.
  • Nazir MU; Department of Respiratory Medicine, The Second Affiliated Hospital of Dalian Medical University, 456 Zhongshan road, 116027, Dalian, Liaoning, People's Republic of China.
  • Lu M; Department of Orthopaedic Surgery, The First Affiliated Hospital of Dalian Medical University, 222 Zhongshan road, 116011, Dalian, Liaoning, People's Republic of China.
BMC Musculoskelet Disord ; 19(1): 73, 2018 03 06.
Article em En | MEDLINE | ID: mdl-29510693
ABSTRACT

BACKGROUND:

Clinically there are different fixation methods used for fixation of the posterior malleolar fractures (PMF), but the best treatment modality is still not clear. Few studies have concentrated on this issue, least of all using a biomechanical comparison. The purpose of this study was to carry out a computational comparative biomechanics of three different commonly used fixation constructs for the fixation of PMF by finite element analysis (FEA).

METHODS:

Computed tomography (CT) images were used to reconstruct three dimensional (3D) model of the tibia. Computer aided design (CAD) software was used to design 3D models of PMF. Finally, 3D models of PMF fixed with two antero-posterior (AP) lag screws, two postero-anterior (PA) lag screws and posterior plate were simulated through computational processing. Simulated loads of 500 N, 1000 N and 1500 N were applied to the PMF and proximal ends of the models were fixed in all degrees of freedom. Output results representing the model von Mises stress, relative fracture micro-motion and vertical displacement of the fracture fragment were analyzed.

RESULTS:

The mean vertical displacement value in the posterior plate group (0.52 mm) was lower than AP (0.68 mm) and PA (0.69 mm) lag groups. Statistically significant low amount of the relative micro-motion (P < 0.05) was observed in the posterior plate group.

CONCLUSIONS:

It was concluded that the posterior plate is biomechanically the most stable fixation method for fixation of PMF.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Placas Ósseas / Parafusos Ósseos / Análise de Elementos Finitos / Imageamento Tridimensional / Fraturas do Tornozelo Limite: Humans Idioma: En Revista: BMC Musculoskelet Disord Assunto da revista: FISIOLOGIA / ORTOPEDIA Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Placas Ósseas / Parafusos Ósseos / Análise de Elementos Finitos / Imageamento Tridimensional / Fraturas do Tornozelo Limite: Humans Idioma: En Revista: BMC Musculoskelet Disord Assunto da revista: FISIOLOGIA / ORTOPEDIA Ano de publicação: 2018 Tipo de documento: Article