Your browser doesn't support javascript.
loading
Ionization of tocopherols and tocotrienols in atmospheric pressure chemical ionization.
Bartosinska, Ewa; Borsuk-De Moor, Agnieszka; Siluk, Danuta; Markuszewski, Michal J; Wiczling, Pawel.
Afiliação
  • Bartosinska E; Department of Biopharmaceutics and Pharmacodynamics, Medical University of Gdansk, Hallera 107, 80-416, Gdansk, Poland.
  • Borsuk-De Moor A; Department of Biopharmaceutics and Pharmacodynamics, Medical University of Gdansk, Hallera 107, 80-416, Gdansk, Poland.
  • Siluk D; Department of Biopharmaceutics and Pharmacodynamics, Medical University of Gdansk, Hallera 107, 80-416, Gdansk, Poland.
  • Markuszewski MJ; Department of Biopharmaceutics and Pharmacodynamics, Medical University of Gdansk, Hallera 107, 80-416, Gdansk, Poland.
  • Wiczling P; Department of Biopharmaceutics and Pharmacodynamics, Medical University of Gdansk, Hallera 107, 80-416, Gdansk, Poland.
Rapid Commun Mass Spectrom ; 32(11): 919-927, 2018 Jun 15.
Article em En | MEDLINE | ID: mdl-29578620
ABSTRACT
RATIONALE Tocopherols and tocotrienols are chemical compounds insusceptible to the ionization process under atmospheric pressure conditions. Therefore, the selection of the optimal ion source settings for their quantification requires special attention. The aim of this study was to analyse the influence of the APCI source parameters on the response of tocochromanols and two related compounds.

METHODS:

Standard solutions of target compounds were injected on the high-performance liquid chromatography/atmospheric pressure chemical ionization tandem mass spectrometry (HPLC/APCI-MS/MS) system separately and analysed with 30 randomly selected ion source settings. The obtained responses were modelled by multivariate linear regression with least absolute shrinkage and selection operator. The developed models were used to choose the best APCI conditions.

RESULTS:

Multivariate linear models were built for eight tocochromanols, trolox and BHT. The APCI settings derived from the models did not increase the peak areas obtained for T and T3 during the ionization process. Ionization conditions based on models for trolox and BHT improved analytical responses by 12-36% and 4-32%, respectively. The application of the ion source settings optimal for trolox and BHT to tocochromanols did not result in better analytical responses.

CONCLUSIONS:

The ionization pattern of tocochromanols in the APCI source is problematic and should be further investigated. Modelling methodology for response improvement presented in this study can be applied in similar studies.

Texto completo: 1 Base de dados: MEDLINE Idioma: En Revista: Rapid Commun Mass Spectrom Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Polônia

Texto completo: 1 Base de dados: MEDLINE Idioma: En Revista: Rapid Commun Mass Spectrom Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Polônia