Quality assurance of geometric accuracy based on an electronic portal imaging device and log data analysis for Dynamic WaveArc irradiation.
Hirashima, Hideaki; Miyabe, Yuki; Nakamura, Mitsuhiro; Mukumoto, Nobutaka; Mizowaki, Takashi; Hiraoka, Masahiro.
Afiliação
  • Hirashima H; Department of Radiation Oncology and Image-applied therapy, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
  • Miyabe Y; Department of Radiation Oncology and Image-applied therapy, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
  • Nakamura M; Department of Radiation Oncology and Image-applied therapy, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
  • Mukumoto N; Department of Radiation Oncology and Image-applied therapy, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
  • Mizowaki T; Department of Radiation Oncology and Image-applied therapy, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
  • Hiraoka M; Department of Radiation Oncology and Image-applied therapy, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
J Appl Clin Med Phys ; 19(3): 234-242, 2018 May.
Article em En | MEDLINE | ID: mdl-29633542
The purpose of this study was to develop a simple verification method for the routine quality assurance (QA) of Dynamic WaveArc (DWA) irradiation using electronic portal imaging device (EPID) images and log data analysis. First, an automatic calibration method utilizing the outermost multileaf collimator (MLC) slits was developed to correct the misalignment between the center of the EPID and the beam axis. Moreover, to verify the detection accuracy of the MLC position according to the EPID images, various positions of the MLC with intentional errors in the range 0.1-1 mm were assessed. Second, to validate the geometric accuracy during DWA irradiation, tests were designed in consideration of three indices. Test 1 evaluated the accuracy of the MLC position. Test 2 assessed dose output consistency with variable dose rate (160-400 MU/min), gantry speed (2.2-6°/s), and ring speed (0.5-2.7°/s). Test 3 validated dose output consistency with variable values of the above parameters plus MLC speed (1.6-4.2 cm/s). All tests were delivered to the EPID and compared with those obtained using a stationary radiation beam with a 0° gantry angle. Irradiation log data were recorded simultaneously. The 0.1-mm intentional error on the MLC position could be detected by the EPID, which is smaller than the EPID pixel size. In Test 1, the MLC slit widths agreed within 0.20 mm of their exposed values. The averaged root-mean-square error (RMSE) of the dose outputs was less than 0.8% in Test 2 and Test 3. Using log data analysis in Test 3, the RMSE between the planned and recorded data was 0.1 mm, 0.12°, and 0.07° for the MLC position, gantry angle, and ring angle, respectively. The proposed method is useful for routine QA of the accuracy of DWA.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Algoritmos / Planejamento da Radioterapia Assistida por Computador / Imagens de Fantasmas / Radioterapia de Intensidade Modulada / Equipamentos e Provisões Elétricas Limite: Humans Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Algoritmos / Planejamento da Radioterapia Assistida por Computador / Imagens de Fantasmas / Radioterapia de Intensidade Modulada / Equipamentos e Provisões Elétricas Limite: Humans Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Japão