Quantification of human C1 esterase inhibitor protein using an automated turbidimetric immunoassay.
Tange, Clare E; Kaur, Amrit; Verma, Nisha; Hickey, Alaco; Grigoriadou, Sofia; Scott, Chris; Kiani, Sorena; Steven, Rachael; Ponsford, Mark; El-Shanawany, Tariq; Jolles, Stephen; Harding, Stephen; Parker, Antony R.
Afiliação
  • Tange CE; The Binding Site Group Limited, Birmingham, UK.
  • Kaur A; The Binding Site Group Limited, Birmingham, UK.
  • Verma N; Barts Health NHS Trust, London, UK.
  • Hickey A; Barts Health NHS Trust, London, UK.
  • Grigoriadou S; Barts Health NHS Trust, London, UK.
  • Scott C; Barts Health NHS Trust, London, UK.
  • Kiani S; Barts Health NHS Trust, London, UK.
  • Steven R; Immunodeficiency Centre for Wales, University Hospital of Wales, Cardiff, UK.
  • Ponsford M; Immunodeficiency Centre for Wales, University Hospital of Wales, Cardiff, UK.
  • El-Shanawany T; Immunodeficiency Centre for Wales, University Hospital of Wales, Cardiff, UK.
  • Jolles S; Immunodeficiency Centre for Wales, University Hospital of Wales, Cardiff, UK.
  • Harding S; The Binding Site Group Limited, Birmingham, UK.
  • Parker AR; The Binding Site Group Limited, Birmingham, UK.
J Clin Lab Anal ; 33(1): e22627, 2019 Jan.
Article em En | MEDLINE | ID: mdl-30058083
ABSTRACT

BACKGROUND:

Impaired levels or function of C1 inhibitor (C1-INH) results in angioedema due to increased bradykinin. It is important to distinguish between angioedema related to C1-INH deficiency and that caused by other mechanisms, as treatment options are different. In hereditary (HAE) and acquired (AAE) angioedema, C1-INH concentration is measured to aid patient diagnosis. Here, we describe an automated turbidimetric assay to measure C1-INH concentration on the Optilite® analyzer.

METHODS:

Linearity, precision, and interference were established over a range of C1-INH concentrations. The 95th percentile reference interval was generated from 120 healthy adult donors. To compare the Optilite C1-INH assay with a predicate assay used in a clinical laboratory, samples sent for C1-INH investigation were used. The predicate results were provided to allow comparison.

RESULTS:

The Optilite C1-INH assay was linear across the measuring range at the standard sample dilution. Intra and interassay variability was <6%. The 95th percentile adult reference interval for the assay was 0.21-0.38 g/L. There was a strong correlation between the Optilite concentrations and those generated with the predicate assay (R2  = 0.94, P < 0.0001, slope y = 0.83x). All patients with Type I HAE (n = 24) and AAE (n = 3) tested had concentrations below the measuring range in both assays, while all patients with unspecified angioedema (UAE), not diagnosed with HAE or AAE had values within the reference range.

CONCLUSION:

The Optilite assay allows the automated and precise quantification of C1-INH concentrations in patient samples. It could therefore be used as a tool to aid the investigation of patients with angioedema.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Proteína Inibidora do Complemento C1 / Imunoturbidimetria Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Clin Lab Anal Assunto da revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Proteína Inibidora do Complemento C1 / Imunoturbidimetria Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Clin Lab Anal Assunto da revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Reino Unido