Comparison of the TOFscan and the TOF-Watch SX during Recovery of Neuromuscular Function.
Murphy, Glenn S; Szokol, Joseph W; Avram, Michael J; Greenberg, Steven B; Shear, Torin D; Deshur, Mark; Benson, Jessica; Newmark, Rebecca L; Maher, Colleen E.
Afiliação
  • Murphy GS; From the Department of Anesthesiology, NorthShore University HealthSystem, University of Chicago Pritzker School of Medicine, Evanston, Illinois (G.S.M., J.W.S., S.B.G., T.D.S., M.D., J.B., R.L.N., C.E.M.) the Department of Anesthesiology, Northwestern University Feinberg School of Medicine, Chicago, Illinois (M.J.A.).
Anesthesiology ; 129(5): 880-888, 2018 11.
Article em En | MEDLINE | ID: mdl-30130260
WHAT WE ALREADY KNOW ABOUT THIS TOPIC: WHAT THIS ARTICLE TELLS US THAT IS NEW: BACKGROUND:: Quantitative neuromuscular monitoring is required to ensure neuromuscular function has recovered completely at the time of tracheal extubation. The TOFscan (Drager Technologies, Canada) is a new three-dimensional acceleromyography device that measures movement of the thumb in multiple planes. The aim of this observational investigation was to assess the agreement between nonnormalized and normalized train-of-four values obtained with the TOF-Watch SX (Organon, Ireland) and those obtained with the TOFscan during recovery from neuromuscular blockade. METHODS: Twenty-five patients were administered rocuronium, and spontaneous recovery of neuromuscular blockade was allowed to occur. The TOFscan and TOF-Watch SX devices were applied to opposite arms. A preload was applied to the TOF-Watch SX, and calibration was performed before rocuronium administration. Both devices were activated, and train-of-four values were obtained every 15 s. Modified Bland-Altman analyses were conducted to compare train-of-four ratios measured with the TOFscan to those measured with the TOF-Watch SX (when train-of-four thresholds of 0.2 to 1.0 were achieved). RESULTS: Bias and 95% limits of agreement between the TOF-Watch SX and the TOFscan at nonnormalized train-of-four ratios between 0.2 and 1.0 were 0.021 and -0.100 to 0.141, respectively. When train-of-four measures with the TOF-Watch SX were normalized, bias and 95% limits of agreement between the TOF-Watch SX and the TOFscan at ratios between 0.2 and 1.0 were 0.015 and -0.097 to 0.126, respectively. CONCLUSIONS: Good agreement between the TOF-Watch SX with calibration and preload application and the uncalibrated TOFscan was observed throughout all stages of neuromuscular recovery.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Período de Recuperação da Anestesia / Bloqueio Neuromuscular / Monitoração Neuromuscular / Acelerometria Tipo de estudo: Observational_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Anesthesiology Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Período de Recuperação da Anestesia / Bloqueio Neuromuscular / Monitoração Neuromuscular / Acelerometria Tipo de estudo: Observational_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Anesthesiology Ano de publicação: 2018 Tipo de documento: Article