High-Grade Dural Arteriovenous Fistulas : Use of Kaneka ED Coils with the Marathon Microcatheter for Transvenous Coil Embolization.
Bhogal, P; AlMatter, M; Hellstern, V; Bäzner, H; Ganslandt, O; Henkes, H; Aguilar Pérez, M.
Afiliação
  • Bhogal P; Neuroradiological Clinic, Neurocenter, Klinikum Stuttgart, Kriegsbergstraße 60, 70174, Stuttgart, Germany. bhogalweb@aol.com.
  • AlMatter M; Neurological Clinic, Neurocenter, Klinikum Stuttgart, Stuttgart, Germany. bhogalweb@aol.com.
  • Hellstern V; Neuroradiological Clinic, Neurocenter, Klinikum Stuttgart, Kriegsbergstraße 60, 70174, Stuttgart, Germany.
  • Bäzner H; Neuroradiological Clinic, Neurocenter, Klinikum Stuttgart, Kriegsbergstraße 60, 70174, Stuttgart, Germany.
  • Ganslandt O; Neurological Clinic, Neurocenter, Klinikum Stuttgart, Stuttgart, Germany.
  • Henkes H; Neurosurgical Clinic, Neurocenter, Klinikum Stuttgart, Stuttgart, Germany.
  • Aguilar Pérez M; Neuroradiological Clinic, Neurocenter, Klinikum Stuttgart, Kriegsbergstraße 60, 70174, Stuttgart, Germany.
Clin Neuroradiol ; 29(4): 653-660, 2019 Dec.
Article em En | MEDLINE | ID: mdl-30167712
BACKGROUND: Dural arteriovenous fistulas (DAVF) are commonly encountered lesions that can be treated both transvenously, transarterially or using a combined approach. OBJECTIVE: Transvenous coil embolization of DAVF is a recognized treatment option but can be challenging. In this context this article presents clinical experience using the Kaneka ED10 ExtraSoft coils in combination with the Marathon microcatheter to treat high grade DAVF. The physical properties of these coils and the microcatheter were also determined. MATERIAL AND METHODS: All patients with high grade DAVF treated with the Marathon and the Kaneka ED COIL ∞10 ExtraSoft coils were retrospectively identified. The clinical presentation, location, grade of the lesion, clinical and radiological follow-up data were recorded. Bench side studies were performed to determine the physical properties of the Marathon catheter in comparison to the SL10 and Headway Duo as well the maximum width of the Kaneka pusher wire in comparison to Hypersoft, Target and Axium Prime coils. RESULTS: A total of 8 patients with 9 DAVF with 3 Cognard 3 and 6 Cognard 4 lesions were identified. All the DAVF's were occluded either at the end of the procedure or on follow-up imaging. On bench side tests the Marathon microcatheter had the most flexible distal tip and distal shaft in comparison to the SL10 and Headway Duo. The proximal shaft of the Marathon was stiffer than the SL10. The Kaneka ED COIL ∞10 ExtraSoft had the smallest distal width and were the only coils tested that could be deployed through a Marathon microcatheter. CONCLUSION: The combination of the Marathon microcatheter and Kaneka ED COIL ∞10 ExtraSoft is useful for the treatment of high grade DAVF.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Malformações Vasculares do Sistema Nervoso Central / Embolização Terapêutica Tipo de estudo: Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Neuroradiol Assunto da revista: NEUROLOGIA / RADIOLOGIA Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Malformações Vasculares do Sistema Nervoso Central / Embolização Terapêutica Tipo de estudo: Observational_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Neuroradiol Assunto da revista: NEUROLOGIA / RADIOLOGIA Ano de publicação: 2019 Tipo de documento: Article País de afiliação: Alemanha