CEST MR-Fingerprinting: Practical considerations and insights for acquisition schedule design and improved reconstruction.
Perlman, Or; Herz, Kai; Zaiss, Moritz; Cohen, Ouri; Rosen, Matthew S; Farrar, Christian T.
Afiliação
  • Perlman O; Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, Charlestown, Massachusetts.
  • Herz K; Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany.
  • Zaiss M; IMPRS for Cognitive and Systems Neuroscience, University of Tübingen, Tübingen, Germany.
  • Cohen O; Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany.
  • Rosen MS; Memorial Sloan Kettering Cancer Center, New York, New York.
  • Farrar CT; Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, Charlestown, Massachusetts.
Magn Reson Med ; 83(2): 462-478, 2020 02.
Article em En | MEDLINE | ID: mdl-31400034
ABSTRACT

PURPOSE:

To understand the influence of various acquisition parameters on the ability of CEST MR-Fingerprinting (MRF) to discriminate different chemical exchange parameters and to provide tools for optimal acquisition schedule design and parameter map reconstruction.

METHODS:

Numerical simulations were conducted using a parallel computing implementation of the Bloch-McConnell equations, examining the effect of TR, TE, flip-angle, water T1 and T2 , saturation-pulse duration, power, and frequency on the discrimination ability of CEST-MRF. A modified Euclidean distance matching metric was evaluated and compared to traditional dot product matching. L-Arginine phantoms of various concentrations and pH were scanned at 4.7T and the results compared to numerical findings.

RESULTS:

Simulations for dot product matching demonstrated that the optimal flip-angle and saturation times are 30∘ and 1100 ms, respectively. The optimal maximal saturation power was 3.4 µT for concentrated solutes with a slow exchange rate, and 5.2 µT for dilute solutes with medium-to-fast exchange rates. Using the Euclidean distance matching metric, much lower maximum saturation powers were required (1.6 and 2.4 µT, respectively), with a slightly longer saturation time (1500 ms) and 90∘ flip-angle. For both matching metrics, the discrimination ability increased with the repetition time. The experimental results were in agreement with simulations, demonstrating that more than a 50% reduction in scan-time can be achieved by Euclidean distance-based matching.

CONCLUSIONS:

Optimization of the CEST-MRF acquisition schedule is critical for obtaining the best exchange parameter accuracy. The use of Euclidean distance-based matching of signal trajectories simultaneously improved the discrimination ability and reduced the scan time and maximal saturation power required.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Imageamento por Ressonância Magnética Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Imageamento por Ressonância Magnética Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2020 Tipo de documento: Article