Your browser doesn't support javascript.
loading
1 H-guided reconstruction of 19 F gas MRI in COPD patients.
Obert, Arnd Jonathan; Gutberlet, Marcel; Kern, Agilo Luitger; Kaireit, Till Frederik; Grimm, Robert; Wacker, Frank; Vogel-Claussen, Jens.
Afiliação
  • Obert AJ; Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany.
  • Gutberlet M; Biomedical Research in Endstage and Obstructive Lung Disease Hannover, German Center for Lung Research, Hannover, Germany.
  • Kern AL; Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany.
  • Kaireit TF; Biomedical Research in Endstage and Obstructive Lung Disease Hannover, German Center for Lung Research, Hannover, Germany.
  • Grimm R; Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany.
  • Wacker F; Biomedical Research in Endstage and Obstructive Lung Disease Hannover, German Center for Lung Research, Hannover, Germany.
  • Vogel-Claussen J; Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany.
Magn Reson Med ; 84(3): 1336-1346, 2020 09.
Article em En | MEDLINE | ID: mdl-32060989
ABSTRACT

PURPOSE:

To reduce acquisition time and improve image quality and robustness of ventilation assessment in a single breath-hold using 1 H-guided reconstruction of fluorinated gas (19 F) MRI.

METHODS:

Reconstructions constraining total variation in the image domain, L1 norm in the wavelet domain, and directional total variation between 19 F and 1 H images were compared in order to accelerate 19 F ventilation imaging using retrospectively undersampled data from a healthy volunteer. Using the optimal constrained reconstruction in 8 patients with chronic obstructive pulmonary disease (16-seconds breath-hold), ventilation maps of various acceleration factors (2-fold to 13-fold) were compared with maps of the full data set using the Dice coefficient, difference in volume defect percentage and overlap percentage, as well as hyperpolarized 129 Xe gas MRI.

RESULTS:

The reconstruction constraining total variation and directional total variation simultaneously performed best in the healthy volunteer (RMS error = 0.07, structural similarity index = 0.77) for a measurement time of 2 seconds. Using the same reconstruction in the patients with chronic obstructive pulmonary disease, the Dice coefficient of defect volumes was 0.86 ± 0.05, the mean difference in volume defect percentage was -1.0 ± 1.7 percentage points, and the overlap percentage was 87% ± 2% for a measurement time of 6 seconds. Between volume defect percentage of 19 F and 129 Xe, a linear correlation (r = 0.75; P = .03) was found, with 19 F volume defect percentage being significantly higher (mean difference = 11%; P = .04).

CONCLUSION:

1 H-guided reconstruction of pulmonary 19 F gas MRI enables reduction of acquisition time while maintaining image quality and robustness of functional parameters.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doença Pulmonar Obstrutiva Crônica Tipo de estudo: Observational_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doença Pulmonar Obstrutiva Crônica Tipo de estudo: Observational_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Alemanha