Your browser doesn't support javascript.
loading
Agreement and repeatability of four different devices to measure non-invasive tear breakup time (NIBUT).
Bandlitz, Stefan; Peter, Barbara; Pflugi, Tanja; Jaeger, Kai; Anwar, Aaisha; Bikhu, Paramdeep; Nosch, Daniela S; Wolffsohn, James S.
Afiliação
  • Bandlitz S; Höhere Fachschule für Augenoptik Köln, Cologne School of Optometry, Cologne, Germany; Optometry and Vision Science, Aston University, Birmingham, UK. Electronic address: bandlitz@hfak.de.
  • Peter B; Institute of Optometry, University of Applied Sciences and Arts Northwestern Switzerland (FHNW), Olten, Switzerland.
  • Pflugi T; Institute of Optometry, University of Applied Sciences and Arts Northwestern Switzerland (FHNW), Olten, Switzerland.
  • Jaeger K; Höhere Fachschule für Augenoptik Köln, Cologne School of Optometry, Cologne, Germany.
  • Anwar A; Optometry and Vision Science, Aston University, Birmingham, UK.
  • Bikhu P; Optometry and Vision Science, Aston University, Birmingham, UK.
  • Nosch DS; Institute of Optometry, University of Applied Sciences and Arts Northwestern Switzerland (FHNW), Olten, Switzerland.
  • Wolffsohn JS; Optometry and Vision Science, Aston University, Birmingham, UK.
Cont Lens Anterior Eye ; 43(5): 507-511, 2020 10.
Article em En | MEDLINE | ID: mdl-32143961
ABSTRACT

PURPOSE:

Since tear film stability can be affected by fluorescein, the Dry Eye Workshop (DEWSII) recommended non-invasive measurement of tear breakup time (NIBUT). The aim of this study was to investigate the agreement and repeatability of four different instruments in the measurement of NIBUT.

METHODS:

72 participants (mean 24.2 ± 3.6 years) were recruited for this multi-centre, cross-sectional study. NIBUT was measured three times from one eye using each of the instruments in randomized order on two separate sessions during a day, separated by at least 2 h. NIBUT was performed at three sites (Switzerland, Germany and UK) using three subjective instruments, Tearscope Plus (Keeler, Windsor, UK) (TS), Polaris (bon Optic, Lübeck, Germany) (POL), EasyTear Viewplus (Easytear, Rovereto, Italy) (ET) and the objective Keratograph 5 M (Oculus Optikgeräte GmbH, Wetzlar, Germany) (KER). As the latter instrument only analyses for 24 s, all data was capped at this value.

RESULTS:

NIBUT measurements (average of both sessions) between the four instruments were not statistically significantly different TS (median 10.4, range 2.0-24.0 s), POL (10.1, 1.0-24.0 s), ET (10.6, 1.0-24.0 s) and KER (11.1, 2.6-24.0 s) (p = 0.949). The objective KER measures were on average (1.2 s ± 9.6 s, 95 % confidence interval) greater than the subjective evaluations of NIBUT with the other instruments (mean difference 0.4 s ± 7.7 s, 95 % confidence interval), resulting in a higher limits of agreement. The slope was -0.08 to 0.11 indicating no bias in the difference between instruments with the magnitude of the NIBUT. Repeated measurements from the two sessions were not significantly different for TS (p = 0.584), POL (p = 0.549), ET (p = 0.701) or KER (p = 0.261).

CONCLUSIONS:

The four instruments evaluated for their measurement of tear stability were reasonably repeatable and give similar average results.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Lágrimas / Síndromes do Olho Seco Tipo de estudo: Clinical_trials / Diagnostic_studies / Observational_studies / Prevalence_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Cont Lens Anterior Eye Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Lágrimas / Síndromes do Olho Seco Tipo de estudo: Clinical_trials / Diagnostic_studies / Observational_studies / Prevalence_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Cont Lens Anterior Eye Ano de publicação: 2020 Tipo de documento: Article