Your browser doesn't support javascript.
loading
Non-contact home-adapted device estimates sleep stages in middle-aged men: A preliminary study.
Yoon, Yeong Sook; Hahm, Jarang; Kim, Kwang Ki; Park, Su Kyung; Oh, Sang Woo.
Afiliação
  • Yoon YS; Department of Family Medicine, Inje University Ilsan Paik Hospital, Juhwa-ro, Ilsanseo-gu, Goyang-Si, Gyeonggi-Do, Korea.
  • Hahm J; Department of Family Medicine, Inje University Ilsan Paik Hospital, Juhwa-ro, Ilsanseo-gu, Goyang-Si, Gyeonggi-Do, Korea.
  • Kim KK; Department of Neurology, Dongguk University Ilsan Hospital, Dongguk-ro, Goyang-si, Ilsandong-gu, Gyeonggi-do, Korea.
  • Park SK; Department of Family Medicine, Inje University Ilsan Paik Hospital, Juhwa-ro, Ilsanseo-gu, Goyang-Si, Gyeonggi-Do, Korea.
  • Oh SW; Department of Neurology, Dongguk University Ilsan Hospital, Dongguk-ro, Goyang-si, Ilsandong-gu, Gyeonggi-do, Korea.
Technol Health Care ; 28(4): 439-446, 2020.
Article em En | MEDLINE | ID: mdl-32444584
ABSTRACT

BACKGROUND:

Sleep monitoring is essential to maintain a healthy life, especially for the elderly who want to age well. Among various forms of sleep devices, the non-wearable and home-adapted device might be preferred because it can be easily used.

OBJECTIVE:

In this study, we evaluated the performance of a non-contact home-adapted device compared to polysomnography (PSG), a gold standard method.

METHODS:

As a preliminary study, eight subjects were recorded over fourteen nights through PSG. The non-contact home-adapted device comprised a microwave sensor, passive infrared sensor, and smartphone application. Through the device, heart rate, respiratory rate, and body movement were collected and used to estimate sleep stages. Sleep structure was labeled in four classes wake, REM, light, and deep sleep, and were classified using a weighted k-nearest neighbor algorithm.

RESULTS:

The device correctly estimated sleep structures with an overall epoch-by-epoch accuracy of 98.65% ± 0.05% based on leave-one-out cross-validation. The device showed significantly positive correlations with PSG in sleep indices including total sleep time, sleep efficiency, and wake after sleep onset.

CONCLUSIONS:

Our findings demonstrate a good performance of this non-contact and home-adapted device and suggest its suitability for sleep monitoring.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fases do Sono Limite: Humans / Male / Middle aged Idioma: En Revista: Technol Health Care Assunto da revista: ENGENHARIA BIOMEDICA / SERVICOS DE SAUDE Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fases do Sono Limite: Humans / Male / Middle aged Idioma: En Revista: Technol Health Care Assunto da revista: ENGENHARIA BIOMEDICA / SERVICOS DE SAUDE Ano de publicação: 2020 Tipo de documento: Article