Development and Validation of a Free Carbamazepine Assay on an Automated Chemistry Analyzer.
Zhang, Xiaochun; Boyert, Nicole; Mitro, Matthew; Tang, Xiaoying; Bowers, Kathleen; McShane, Adam J.
Afiliação
  • Zhang X; Pathology and Laboratory Medicine Institute, Cleveland Clinic, Cleveland, OH.
  • Boyert N; Pathology and Laboratory Medicine Institute, Cleveland Clinic, Cleveland, OH.
  • Mitro M; Pathology and Laboratory Medicine Institute, Cleveland Clinic, Cleveland, OH.
  • Tang X; Pathology and Laboratory Medicine Institute, Cleveland Clinic, Cleveland, OH.
  • Bowers K; Pathology and Laboratory Medicine Institute, Cleveland Clinic, Cleveland, OH.
  • McShane AJ; Pathology and Laboratory Medicine Institute, Cleveland Clinic, Cleveland, OH.
J Appl Lab Med ; 5(2): 357-362, 2020 03 01.
Article em En | MEDLINE | ID: mdl-32445379
BACKGROUND: Carbamazepine is an effective drug for treating seizures and trigeminal neuralgia. Therapeutic drug monitoring of free carbamazepine in serum can be useful in situations that drug--protein binding is altered to guide regimen adjustment and to aid in the diagnosis of clinical toxicity. METHODS: Separation of the nonprotein bound carbamazepine was achieved via ultrafiltration through a molecular weight cut-off filter. A method for free carbamazepine measurement was developed on the automated cobas chemistry analyzers (Roche Diagnostics) by modifying the Carbamazepine Gen 4 assay (Roche Diagnostics). Assay performance characteristics were established including precision, accuracy, reportable range, analytical specificity, and stability. RESULTS: The intra- and inter-assay imprecision was 0%-1.4% and 2.4%-5.1%, respectively. The lower limit of quantitation was 0.3 µg/mL, and the assay was linear up to 10.0 µg/mL. A spike recovery study, using reference standard material, showed recovery was 93.5%--101.3% across the analytical measurement range. Method comparison with a reference laboratory method demonstrated equivalent performance with a slope of 1.01, intercept of 0.09, and correlation coefficient of 0.9948. CONCLUSION: This assay provides a simple and accurate method for monitoring free carbamazepine with a fast turnaround time.