Dynamic locking screws in proximal humeral plate osteosynthesis demonstrate superior fixation properties: a biomechanical study.
Sandmann, Gunther; Ateschrang, Atesch; Freude, Thomas; Stöckle, Ulrich; Schmölz, Werner; Konrads, Christian; Döbele, Stefan.
Afiliação
  • Sandmann G; Sportsclinic Ravensburg, Ravensburg, Germany.
  • Ateschrang A; Department of Trauma and Reconstructive Surgery, BG Klinik, University of Tübingen, Schnarrenbergstr. 95, 72076, Tübingen, Germany.
  • Freude T; Department of Orthopaedics and Traumatology, University of Salzburg, Salzburg, Austria.
  • Stöckle U; Center for Musculoskeletal Surgery, Charité - University Medical Center Berlin, Berlin, Germany.
  • Schmölz W; Department of Biomechanics, University of Innsbruck, Innsbruck, Austria.
  • Konrads C; Department of Trauma and Reconstructive Surgery, BG Klinik, University of Tübingen, Schnarrenbergstr. 95, 72076, Tübingen, Germany. christian.konrads@gmail.com.
  • Döbele S; Department of Trauma and Reconstructive Surgery, BG Klinik, University of Tübingen, Schnarrenbergstr. 95, 72076, Tübingen, Germany.
J Exp Orthop ; 7(1): 82, 2020 Oct 12.
Article em En | MEDLINE | ID: mdl-33044584
ABSTRACT

PURPOSE:

Angular stable implants reduced the complication rate in the treatment of humeral head fractures. But the failure rate is still high. To further reduce the risk of cut-out, cement augmentation of screws was introduced. A reason for failure of plate osteosynthesis might be the extremely high stiffness of the screw-plate interface leading to a loss of reduction and cut-out of screws. A more homogeneous distribution of the forces on all screws may avoid secondary dislocation. We hypothesize that dynamic osteosynthesis minimizes screw loosening and results in a higher load to failure than standard locking screws.

METHODS:

Twelve paired human humerus specimens were analysed. A standardized three-part fracture model with a metaphyseal defect was simulated. Within each pair of humeri, one was fixed with a Philos plate and standard locking screws (LS), whereas the other humerus was fixed with a Philos plate and dynamic locking screws (DLS). A cyclic varus-bending test or a rotation test with increasing loading force was performed until failure of the screw-bone-fixation.

RESULTS:

In the varus bending test, pairs failed by screw loosening in the humeral head. The LS-group reached 2901 (601-5201) load cycles until failure, while the DLS-group failed after 3731 (2001-5601) cycles. This corresponds to a median loading of 195 N for the LS-group and 235 N for the DLS-group (p = 0.028). In the rotation test the LS-group reached a median of 1101 (501-1501) load cycles until failure of fixation occurred, while the DLS-group failed after 1401 (401-2201) cycles (p = 0.225).

CONCLUSIONS:

Plate fixation using dynamic locking screws for the treatment of proximal humerus fractures demonstrated more load cycles until failure compared to standard locking plate osteosynthesis.