Virtual touch imaging quantification shear-wave elastography for breast lesions: the diagnostic value of qualitative and quantitative features.
Luo, T; Zhang, J W; Zhu, Y; Jia, X H; Dong, Y J; Zhan, W W; Zhou, J Q.
Afiliação
  • Luo T; Department of Ultrasound, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai 200025, China.
  • Zhang JW; Department of Ultrasound, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai 200025, China.
  • Zhu Y; Department of Ultrasound, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai 200025, China.
  • Jia XH; Department of Ultrasound, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai 200025, China.
  • Dong YJ; Department of Ultrasound, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai 200025, China. Electronic address: dongyiyin@126.com.
  • Zhan WW; Department of Ultrasound, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai 200025, China.
  • Zhou JQ; Department of Ultrasound, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai 200025, China. Electronic address: zhousu30@126.com.
Clin Radiol ; 76(4): 316.e1-316.e8, 2021 Apr.
Article em En | MEDLINE | ID: mdl-33328105
ABSTRACT

AIM:

To investigate the value of the qualitative and quantitative features of Virtual Touch imaging quantification (VTIQ) shear-wave elastography in the characterisation of breast lesions. MATERIALS AND

METHODS:

Conventional ultrasound (US) and VTIQ were performed in 148 solid breast lesions in 148 women. During qualitative analysis, patterns of VTIQ were categorised into two patterns, 1 and 2. During quantitative analysis, the mean SWV (SWVmean) and the maximum SWV (SWVmax) of each lesion were used. The sensitivity, specificity, and the areas under the receiver operating characteristic (ROC) curve (Az value) were calculated for conventional US, VTIQ, and combined conventional US and VTIQ.

RESULTS:

Malignant lesions were more likely to show VTIQ pattern 2 than the benign lesions (p<0.001). There was no significant difference in the Az values between SWVmean (0.907) and SWVmax (0.902; p=0.572). There was no significant difference in the Az values between the VTIQ pattern (0.884) and SWVmax (p=0.572). The combined conventional US and VTIQ pattern carried a similar Az value (0.949) as compared with the combined conventional US and SWVmax, which yielded an Az value of 0.952 (p=0.683).

CONCLUSION:

The combination of either VTIQ pattern or SWVmax and conventional US may be helpful in the characterisation of benign and malignant breast lesions.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doenças Mamárias / Neoplasias da Mama / Ultrassonografia Mamária / Técnicas de Imagem por Elasticidade Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies / Qualitative_research Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Middle aged Idioma: En Revista: Clin Radiol Ano de publicação: 2021 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doenças Mamárias / Neoplasias da Mama / Ultrassonografia Mamária / Técnicas de Imagem por Elasticidade Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies / Qualitative_research Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Middle aged Idioma: En Revista: Clin Radiol Ano de publicação: 2021 Tipo de documento: Article País de afiliação: China