Rapid, B1 -insensitive, dual-band quasi-adiabatic saturation transfer with optimal control for complete quantification of myocardial ATP flux.
Miller, Jack J; Valkovic, Ladislav; Kerr, Matthew; Timm, Kerstin N; Watson, William D; Lau, Justin Y C; Tyler, Andrew; Rodgers, Christopher; Bottomley, Paul A; Heather, Lisa C; Tyler, Damian J.
Afiliação
  • Miller JJ; Department of Physics, University of Oxford, Oxford, UK.
  • Valkovic L; Department of Physiology, Anatomy and Genetics, University of Oxford, Oxford, UK.
  • Kerr M; Oxford Centre for Clinical Magnetic Resonance Research, John Radcliffe Hospital, Headington, Oxford, UK.
  • Timm KN; Health, Aarhus University, Aarhus, Denmark.
  • Watson WD; Oxford Centre for Clinical Magnetic Resonance Research, John Radcliffe Hospital, Headington, Oxford, UK.
  • Lau JYC; Department of Imaging Methods, Institute of Measurement Science, Slovak Academy of Sciences, Bratislava, Slovakia.
  • Tyler A; Department of Physiology, Anatomy and Genetics, University of Oxford, Oxford, UK.
  • Rodgers C; Department of Physiology, Anatomy and Genetics, University of Oxford, Oxford, UK.
  • Bottomley PA; Oxford Centre for Clinical Magnetic Resonance Research, John Radcliffe Hospital, Headington, Oxford, UK.
  • Heather LC; Department of Physiology, Anatomy and Genetics, University of Oxford, Oxford, UK.
  • Tyler DJ; Oxford Centre for Clinical Magnetic Resonance Research, John Radcliffe Hospital, Headington, Oxford, UK.
Magn Reson Med ; 85(6): 2978-2991, 2021 06.
Article em En | MEDLINE | ID: mdl-33538063
ABSTRACT

PURPOSE:

Phosphorus saturation-transfer experiments can quantify metabolic fluxes noninvasively. Typically, the forward flux through the creatine kinase reaction is investigated by observing the decrease in phosphocreatine (PCr) after saturation of γ-ATP. The quantification of total ATP utilization is currently underexplored, as it requires simultaneous saturation of inorganic phosphate ( Pi ) and PCr. This is challenging, as currently available saturation pulses reduce the already-low γ-ATP signal present.

METHODS:

Using a hybrid optimal-control and Shinnar-Le Roux method, a quasi-adiabatic RF pulse was designed for the dual saturation of PCr and Pi to enable determination of total ATP utilization. The pulses were evaluated in Bloch equation simulations, compared with a conventional hard-cosine DANTE saturation sequence, before being applied to perfused rat hearts at 11.7 T.

RESULTS:

The quasi-adiabatic pulse was insensitive to a >2.5-fold variation in B1 , producing equivalent saturation with a 53% reduction in delivered pulse power and a 33-fold reduction in spillover at the minimum effective B1 . This enabled the complete quantification of the synthesis and degradation fluxes for ATP in 30-45 minutes in the perfused rat heart. While the net synthesis flux (4.24 ± 0.8 mM/s, SEM) was not significantly different from degradation flux (6.88 ± 2 mM/s, P = .06) and both measures are consistent with prior work, nonlinear error analysis highlights uncertainties in the Pi -to-ATP measurement that may explain a trend suggesting a possible imbalance.

CONCLUSIONS:

This work demonstrates a novel quasi-adiabatic dual-saturation RF pulse with significantly improved performance that can be used to measure ATP turnover in the heart in vivo.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Trifosfato de Adenosina / Miocárdio Limite: Animals Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Trifosfato de Adenosina / Miocárdio Limite: Animals Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Reino Unido