Your browser doesn't support javascript.
loading
False Positives in Thyroglobulin Determinations Due to the Presence of Heterophile Antibodies: An Underrecognized and Consequential Clinical Problem.
Barbesino, Giuseppe; Algeciras-Schimnich, Alicia; Bornhorst, Joshua A.
Afiliação
  • Barbesino G; Thyroid Unit, Massachusetts General Hospital-Harvard Medical School, Boston, Massachusetts. Electronic address: gbarbesino@mgh.harvard.edu.
  • Algeciras-Schimnich A; Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, Minnesota.
  • Bornhorst JA; Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, Minnesota.
Endocr Pract ; 27(5): 396-400, 2021 May.
Article em En | MEDLINE | ID: mdl-33581326
ABSTRACT

OBJECTIVE:

To report a case series of thyroid cancer patients in whom false positive results in immunometric assays for thyroglobulin (TgIMA) were caused by heterophilic antibody interference, describe the clinical scenario in which this interference should be suspected, and recommend methods to demonstrate the interference.

METHODS:

Three patients with unexpectedly elevated thyroglobulin results (range, 1.6-75 ng/mL) were studied. In the first patient, thyroglobulin was elevated despite the presence of Tg antibody. In the second patient, suppressed thyroglobulin was higher than a recent stimulated thyroglobulin. In the third patient, thyroglobulin became detectable years after treatment and did not change after thyroid-stimulating hormone stimulation. TgIMA concentration determination was compared to determination by a mass spectrometry method (TgMS). Thyroglobulin was also remeasured after preabsorption with heterophile antibody blocking reagents and after serial dilutions.

RESULTS:

In all cases, thyroglobulin was undetectable by TgMS. In 2 of 3 patients, dilutions provided nonlinear thyroglobulin results. After blocking agent preabsorption, thyroglobulin dropped by 35%, 45%, and 91% in the 3 samples.

CONCLUSION:

False positive thyroglobulin concentrations from heterophilic antibody interference have significant impact on the management of thyroid cancer. Here we show that TgMS assays can be used to rule out heterophilic antibody interference. This interference should be suspected when a detectable thyroglobulin by TgIMA does not respond to thyroid-stimulating hormone or is discordant from the clinical assessment.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Tireoglobulina / Neoplasias da Glândula Tireoide Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Endocr Pract Assunto da revista: ENDOCRINOLOGIA Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Tireoglobulina / Neoplasias da Glândula Tireoide Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Endocr Pract Assunto da revista: ENDOCRINOLOGIA Ano de publicação: 2021 Tipo de documento: Article