Your browser doesn't support javascript.
loading
A comparison of liver fat fraction measurement on MRI at 3T and 1.5T.
Athithan, Lavanya; Gulsin, Gaurav S; House, Michael J; Pang, Wenjie; Brady, Emer M; Wormleighton, Joanne; Parke, Kelly S; Graham-Brown, Matthew; St Pierre, Tim G; Levelt, Eylem; McCann, Gerry P.
Afiliação
  • Athithan L; Department of Cardiovascular Sciences, University of Leicester and NIHR Leicester Cardiovascular Biomedical Research Centre, Glenfield Hospital, Leicester, United Kingdom.
  • Gulsin GS; Department of Cardiovascular Sciences, University of Leicester and NIHR Leicester Cardiovascular Biomedical Research Centre, Glenfield Hospital, Leicester, United Kingdom.
  • House MJ; Department of Physics, The University of Western Australia, Crawley, Western Australia, Australia.
  • Pang W; Resonance Health Ltd, Burswood, Western Australia, Australia.
  • Brady EM; Resonance Health Ltd, Burswood, Western Australia, Australia.
  • Wormleighton J; Department of Cardiovascular Sciences, University of Leicester and NIHR Leicester Cardiovascular Biomedical Research Centre, Glenfield Hospital, Leicester, United Kingdom.
  • Parke KS; Department of Cardiovascular Sciences, University of Leicester and NIHR Leicester Cardiovascular Biomedical Research Centre, Glenfield Hospital, Leicester, United Kingdom.
  • Graham-Brown M; Department of Cardiovascular Sciences, University of Leicester and NIHR Leicester Cardiovascular Biomedical Research Centre, Glenfield Hospital, Leicester, United Kingdom.
  • St Pierre TG; Department of Cardiovascular Sciences, University of Leicester and NIHR Leicester Cardiovascular Biomedical Research Centre, Glenfield Hospital, Leicester, United Kingdom.
  • Levelt E; Department of Physics, The University of Western Australia, Crawley, Western Australia, Australia.
  • McCann GP; Department of Cardiovascular Sciences, University of Leicester and NIHR Leicester Cardiovascular Biomedical Research Centre, Glenfield Hospital, Leicester, United Kingdom.
PLoS One ; 16(7): e0252928, 2021.
Article em En | MEDLINE | ID: mdl-34255778
ABSTRACT

PURPOSE:

Volumetric liver fat fraction (VLFF) measurements were made using the HepaFat-Scan® technique at 1.5T and 3T to determine the agreement between the measurements obtained at the two fields.

METHODS:

Sixty patients with type 2 diabetes (67% male, mean age 50.92 ± 6.56yrs) and thirty healthy volunteers (50% male, mean age 48.63 ± 6.32yrs) were scanned on 1.5T Aera and 3T Skyra (Siemens, Erlangen, Germany) MRI scanners on the same day using the HepaFat-Scan® gradient echo protocol with modification of echo times for 3T (TEs 2.38, 4.76, 7.14 ms at 1.5T and 1.2, 2.4, 3.6 ms at 3T). The 3T analyses were performed independently of the 1.5T analyses by a different analyst, blinded from the 1.5T results. Data were analysed for agreement and bias using Bland-Altman methods and intraclass correlation coefficients (ICC). A second cohort of 17 participants underwent interstudy repeatability assessment of VLFF measured by HepaFat-Scan® at 3T.

RESULTS:

A small, but statistically significant mean bias of 0.48% was observed between 3T and 1.5T with 95% limits of agreement -2.2% to 3.2% VLFF. The ICC for agreement between field strengths was 0.983 (95% CI 0.972-0.989). In the repeatability cohort studied at 3T the repeatability coefficient was 4.2%. The ICC for agreement was 0.971 (95% CI 0.921-0.989).

CONCLUSION:

There is minimal bias and excellent agreement between the measures of VLFF using the HepaFat-Scan® at 1.5 and 3T. The test retest repeatability coefficient at 3T is comparable to the 95% limits of agreement between 1.5T and 3T suggesting that measurements can be made interchangeably between field strengths.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Tecido Adiposo / Fígado Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: PLoS One Assunto da revista: CIENCIA / MEDICINA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Tecido Adiposo / Fígado Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: PLoS One Assunto da revista: CIENCIA / MEDICINA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Reino Unido