Validation of Image Qualities of a Novel Four-Mice Bed PET System as an Oncological and Neurological Analysis Tool.
Kang, Kyung Jun; Oh, Se Jong; Nam, Kyung Rok; Ahn, Heesu; Park, Ji-Ae; Lee, Kyo Chul; Choi, Jae Yong.
Afiliação
  • Kang KJ; Division of Applied RI, Korea Institute of Radiological and Medical Sciences, Seoul 01812, Korea.
  • Oh SJ; Division of Applied RI, Korea Institute of Radiological and Medical Sciences, Seoul 01812, Korea.
  • Nam KR; Division of Applied RI, Korea Institute of Radiological and Medical Sciences, Seoul 01812, Korea.
  • Ahn H; Division of Applied RI, Korea Institute of Radiological and Medical Sciences, Seoul 01812, Korea.
  • Park JA; Radiological and Medico-Oncological Sciences, University of science and technology (UST), Seoul 01812, Korea.
  • Lee KC; Division of Applied RI, Korea Institute of Radiological and Medical Sciences, Seoul 01812, Korea.
  • Choi JY; Radiological and Medico-Oncological Sciences, University of science and technology (UST), Seoul 01812, Korea.
J Imaging ; 7(3)2021 Feb 26.
Article em En | MEDLINE | ID: mdl-34460699
ABSTRACT

Background:

Micro-positron emission tomography (micro-PET), a small-animal dedicated PET system, is used in biomedical studies and has the quantitative imaging capabilities of radiotracers. A single-bed system, commonly used in micro-PET, is laborious to use in large-scale studies. Here, we evaluated the image qualities of a multi-bed system.

Methods:

Phantom imaging studies were performed to assess the recovery coefficients (RCs), uniformity, and spill-over ratios (SORs) in water- and air-filled chambers. 18F-FDG and 18F-FPEB PET images of xenograft and normal mice from the multi-bed and single-bed systems were compared.

Results:

For small diameters (< 3 mm), the RC values between the two systems differed significantly. However, for large diameters (> 4 mm), there were no differences in RC values between the two systems. Uniformity and SORs of both systems were within the tolerance limit of 15%. In the oncological study, the estimation of 18F-FDG uptake in the tumor was significantly lower in the multi-bed system than that in the single-bed system. However, 18F-FDG PET in xenograft mice with tumor size > 4 mm revealed the variation between subjects within the multi-bed system group to be less than 12%. In the neurological study, SUV for the multi-bed group was 25-26% lower than that for the single-bed group; however, inter-object variations within the multi-bed system were below 7%.

Conclusions:

Although the multi-bed system showed lower estimation of radiotracer uptake than that of the single-bed system, the inter-subject variations were within acceptable limits. Our results indicate that the multi-bed system can be used in oncological and neurological studies.